Introduction
Approved by the Food and Drug Administration in 1996, the bispectral index (BIS) is the most widely used noninvasive method to estimate level of consciousness [
1,
2]. The BIS is a complex electroencephalography (EEG) parameter that converts raw EEG data from the frontal cortex into a single number between 100 (fully awake) and 0 (isoelectric EEG) [
3].
The BIS sensor is a single adhesive strip placed across the forehead. This position can be problematic in certain surgical procedures involving a fronto-temporal approach, such as those used in plastic surgery and neurosurgery. Although several alternative sensor positions have been examined in previous studies, there remains significant disagreement over their reliability. In one study, occipital placement had a strong relationship (r
2 = 0.961) with standard frontal placement during general anesthesia for neurosurgery [
4]. In contrast, other studies found poor agreement between occipital and frontal placement with a discrepancy of over ± 10 BIS units [
5,
6]. A third study found BIS values with nasal placement had adequate agreement with standard frontal placement [
6].
We hypothesized that BIS values from mandibular sensor placement would be significantly correlated with those from the standard frontal position. To test this hypothesis, we compared BIS values between mandibular and frontal sensor positions in individual patients at different stages of anesthesia.
Materials and Methods
We enrolled 58 patients scheduled for various surgical procedures not involving the head or neck requiring routine general anesthesia. All participants provided written informed consent. Participants were between 18 and 75 years of age and had an American Society of Anesthesiologists (ASA) of I-II. We excluded patients with a history of disabling central nervous or cerebrovascular disease, those currently taking central nervous system active drugs or psychiatric medication, and those who had previously undergone neurosurgery. The Institutional Review Board approved this study.
Before the induction of anesthesia, we applied 2 BIS sensors (BIS™ Quatro Sensors, Aspect Medical Systems, Newton, MA, USA) to each patient, 1 across the forehead and 1 along the mandible, and attached each sensor to an each BIS monitor (BIS-Vista™ monitors, Aspect Medical Systems, Newton, MA, USA). Frontal sensors were applied with circle 1 at the center of the forehead; circle 2, 2.8 cm lateral to circle 1; and circle 3 on either temporal area between the corner of the eye and the hairline. Mandibular sensors were applied to the same side of the face, with circle 1 at the center of the mandible, circle 4 adjacent to the lateral side of lip, and circle 3 at the ipsilateral temporal area (
Fig. 1). Anesthesia was induced with 2 mg/kg of propofol and 0.7 mg/kg of rocuronium was administered as a muscle relaxant. Then Anesthesia was maintained with sevoflurane or desflurane in 50% N
2O. Minimum alveolar concentration range for the inhaled anesthetics was 0.8-1.5. At the completion of surgery, 10 mg of pyridostigmine with 0.4 mg of glycopyrrolate was given to antagonize any residual neuromuscular block. From each BIS monitor, we collected 3 pieces of data at each of 6 times points: BIS score, signal quality index (SQI), and electromyography (EMG) score before the induction of anesthesia (awake), at loss of the eyelash reflex (LOC), after intubation (intubation), after the first surgical incision (incision), every 30 minutes during the intraoperative period (maintenance), and at spontaneous eye opening upon emergence from anesthesia (emergence). Data were downloaded from BIS monitors after each procedure using a USB port for subsequent analysis. Sudden, high EMG scores were identified as artifacts and the associated BIS values were eliminated in data analysis. During induction, patients were instructed to keep their faces relaxed (eyes closed, mouth closed, no facial expressions). In our previous pilot study, we found that the frontal BIS scores fell more quickly than mandibular scores during induction. Therefore, we recorded the difference in time required for the BIS value to reach its lowest level between sensor positions during induction. We complied data from 10 pilot patients in whom the mean BIS values during anesthesia maintenance were 50.1 ± 10.5 (frontal position) and 54.2 ± 5.1 (mandibular position). Given this preliminary data, to achieve an alpha of 0.05 and a power of 80%, we required a sample size of 52 patients. We estimated a 10% drop-out rate, and thus included 58 patients in our study.
Patient data are presented as mean ± SD. We compared BIS scores from both sensor positions using linear regression analysis and used Bland-Altman plots to assess agreement between frontal and mandibular sensors with a 95% confidence interval (CI). The "limit of agreement" was defined as a bias of ± 1.96 SD, in which 95% of the difference between the 2 placements were expected to lie between the limits, which would suggest that the 2 positions could be used interchangeably. According to previous studies [
5,
6], we considered a clinically acceptable level of "limit of agreement" to be ± 10 BIS units. We used scatter plots to evaluate BIS score correlations and calculated the R coefficient. We performed paired t-tests to compare differences in EMG and SQI scores. Statistical analysis was performed using SPSS software (SPSS version 18.0, IBM, USA) and Medcalc for Windows, version 12.5 (MedCalc Software, Ostend, Belgium). We used a P value less than 0.05 as the cutoff for statistical significance.
Results
Patient characteristics are summarized in
Table 1. There were significant correlations between frontal and mandibular BIS values at all time points except emergence (P = 0.077;
Table 2,
Fig. 2). Bland-Altman analysis of frontal and mandibular BIS values resulted in a bias of -3.2 with limits of agreement of 16.5 and -22.9, respectively. Limits of agreement ranged from 9.0 to -12.5 during maintenance and 14.9 to -58.3 at LOC (
Table 3,
Fig. 3).
EMG and SQI varied by time point. EMG scores were lowest intraoperatively and highest during emergence. Mean EMG scores were below 40 at all time points except during awake and emergence time points. Mean EMG scores were significantly different between frontal and mandibular positions during the awake, intubation, and emergence time points: mandibular scores were significantly higher than frontal scores at all time points except LOC, incision and maintenance. Average SQI was over 80. In the frontal position, mean SQI scores were below 80 during awake or intubation time points, while in the mandibular position, they were below 80 during awake, LOC, intubation and emergence time points. Conversely, mean SQI scores were significantly higher in the frontal than in the mandibular position except during incision and maintenance time points (
Table 4).
The time it took for BIS scores to fall to the lowest level during induction varied greatly between frontal and mandibular positions (range 0-120 sec, mean ± SD = 26.9 ± 27.4 sec).
Discussion
Our research shows that BIS, EMG, and SQI scores from the 2 sensor positions vary by anesthesia time point. The correlation coefficient between the 2 positions indicated a strong correlation during intubation (R = 0.845), incision (R = 0.601), and the intraoperative maintenance period (R = 0.607), as well as a weak correlation during awake (R = 0.391) and LOC (R = 0.341) time points. During the emergence time point, there was no correlation between BIS values from the frontal and mandibular positions. We hypothesize that this pattern of correlation is related to physiologic changes that occur in different anesthesia stages: during a stable sedation state with sufficient muscle relaxation and low BIS scores, the correlation coefficient is higher than in other stages.
Bland Altman analysis also reveals similar findings. The mean difference was lowest at incision (-1.0) and highest at loss of consciousness (-21.7), and was relatively lower at intubation (-1.9) and during maintenance (-1.8). The percentage of BIS scores outside of the limit of agreement is only 3.7% overall, and not over 5% in any anesthetic stage. However, there is variability in 95% CI depending on various anesthetic states. Although a clinically acceptable limit of agreement has not been defined, some authors consider it to be ± 10 BIS units [
5,
6]. In our case, 95% CI is over 10 BIS units. However, in our study, we found a relatively narrow limit of agreement during stable states (awake, 14.1/-8.3; and maintenance, 9.0/-12.5) by comparison with the other anesthetic states that were expected a relatively wider change in BIS score (LOC, 14.9/-58.3, and emergence, 14.9/-18.8). These data suggest that sudden changes in BIS score reduce the reliability of mandibular sensor placement.
Overall, BIS values do not agree between the standard frontal position and the alternative mandibular position. But correlation coefficient between two sensor position showed strong relationship (R = 0.607) and Bland Altman analysis revealed low mean difference (-1.8) and relatively low limit of agreement (9.0/-12.5) with only 4.3% of the values lying outside of the limit of agreement during anesthetic maintenance period. On the basis of the clinical values and correlation, it can be clinically acceptable. Therefore, during the anesthesia maintenance period, the mandibular position can be availably used as an alternative position if the operative field renders the standard position unavailable.
EMG scores varied greatly depending on anesthetic state. The difference in mean EMG score was statistically significant between frontal and mandibular sensor positions during the awake and emergence time points. This may be due to increased muscle tone during these stages. Overall, the mean EMG score was higher in the mandibular than in the frontal position. This may be due to more muscle movement in the chin than in the forehead, since this difference became statistically insignificant after administration of the muscle relaxant. The significant difference between mean EMG scores for the 2 sensor positions during intubation was likely due to the jaw manipulation required for endotracheal intubation although sufficient muscle relaxation enough to intubation was exist. Facial muscle EMG activity can produce artifact in the EEG, greatly impacting the BIS score [
7]. Higher EMG in mandible sensor could influence not only relatively higher BIS values of mandible sensor than that of frontal sensor but also negative aspect of equivalent of two sensors. Consistent with this, overall, BIS scores from the mandibular sensor position were higher than those from the frontal position.
Mean SQI scores were relatively lower in the mandibular than in the frontal sensor position. However, during incision and maintenance periods, mean SQI was over 80, and without statistically significant difference, in both sensor positions. SQI measures the quality of the acquired EEG signal (0-100%). SQI is more than 80, it ensured good quality data [
8]. This result indicates the reliability of the mandibular sensor position during the stable anesthetic state. Conversely, mean SQI was significantly higher in the frontal than in the mandibular sensor position during awake, LOC, intubation, and emergence time points, indicating low clinical accuracy of the mandibular position. Other factors such as a strong carotid pulse, use of the upper body forced air warming units or electrocardiography can affect the BIS scores of mandibular sensor as an artifact. And these interferences can lead to false interpretation of BIS readings without any sign of artifact recognition [
9,
10].
After intravenous anesthetics were given, BIS values fell acutely at the LOC time point. In our previous pilot study, we found the frontal BIS values fell more quickly than mandibular BIS values during induction. The time it took for BIS scores to fall to the lowest level during induction varied greatly between frontal and mandibular positions (range, 0-120 sec). This finding suggests that the spread of EEG current to the chin may delay conduction time.
Other authors have proposed that BIS score is a topographic dependent variable in light of the heterogeneous EEG findings in BIS sensors placed on non-frontal areas. One study on occipital placement showed a +10 BIS score bias under deep anesthesia and a -10 BIS bias before induction. Although the nature of the BIS algorithm is not disclosed by the creators of this technology, this result may be due to the predominance of posterior alpha (α) waves in the awake brain and generation of delta (δ) and theta (θ) activity under deep propofol anesthesia [
5]. In contrast, strong correlations have been reported between frontal and occipital sensor placement, as well as between fronto-central and bi-frontal placement [
4,
11]. Unlike the frontal or occipital area, no EEG is generated under the mandible, and thus detectable EEG or BIS is likely conducted from other parts of cerebral cortex. The exact locations remain unknown and warrant further study.
There are several limitations to our study. First, our study did not standardize the anesthetic methods used, which may have influenced the results [
12]. Second, BIS-vista does not generate raw EEG tracing for analysis, we were unable to confirm that the actual EEG recordings were identical. Although the BIS algorithm has not been formally validated, actual EEG recordings at each electrode on frontal and mandibular positions could help to confirm the accuracy and characteristics of the EEG signal arising from the mandible.
In conclusion, during the anesthesia maintenance period, the mandibular position can be availably used as an alternative position if the operative field renders the standard position unavailable.