Korean Journal of Anesthesiology 2006;50(5):506-510.
DOI: https://doi.org/10.4097/kjae.2006.50.5.506   
Assessment of Blood Glucose Values Measured by Blood Gas Analyzer or Portable Glucometer.
Hue Jung Park, Chul Soo Park, Chong Min Park, Keon Hee Ryu, Hae Wone Chang, Eun Jeong Cho, Yoon Ki Lee
Department of Anesthesiology and Pain Medicine, Kangnam St. Mary's Hospital, College of Medicine, The Catholic University of Korea, Seoul, Korea. yklee@catholic.ac.kr
Abstract
BACKGROUND
A portable glucometer is commonly used to immediately check the blood glucose level. In the anesthetic field, some blood gas analyzers can also give a rapid indication of the blood sugar level but the accuracy is unknown. Therefore, this study assessed the accuracy of the blood glucose values measured by either a blood gas analyzer or portable glucometer.
METHODS
Venous blood from diabetic patients was used to measure the glucose level with either a blood gas analyzer or a portable glucometer. The difference and 5% deviation from reference values was analyzed. These values were also assessed using a Bland-Altman plot and clinical significance was examined using a Clarke error grid.
RESULTS
The differences from the reference values were smaller using the blood gas analyzer (1.3 +/- 7.8 mg/dl) than using the portable glucometer (-5.1 +/- 16.7 mg/dl)(P < 0.01). 73.4% of the values measured by the blood gas analyzer and 40.0% of those measured by the portable glucometer were within 5% of the reference value. The 95% limits of agreement in the difference ranged from -14.3 to 16.9 in the blood gas analyzer and -38.5 to 28.2 in the portable glucometer. Error grid analysis showed that 100% of the values measured by the blood gas analyzer were located in zone A. When locating the values measured using the portable glucometer, 95.6% were located in zone A, and the remaining 4.4% are located in zone B.
CONCLUSIONS
The blood gas analyzer measures the blood glucose more accurately than the portable glucometer. However, the blood glucose values measured by the portable glucometer are clinically acceptable.
Key Words: bland altman analysis; blood gas analyzer; error grid analysis; glucose; portable glucometer
TOOLS
Share :
Facebook Twitter Linked In Line it
METRICS Graph View
  • 2 Crossref
  •    
  • 1,935 View
  • 28 Download


ABOUT
ARTICLE CATEGORY

Browse all articles >

BROWSE ARTICLES
AUTHOR INFORMATION
Editorial Office
101-3503, Lotte Castle President, 109 Mapo-daero, Mapo-gu, Seoul 04146, Korea
(서울특별시 마포구 마포대로 109 롯데캐슬 프레지던트 101동 3503호)
Tel: +82-2-792-5128    Fax: +82-2-792-4089    E-mail: journal@anesthesia.or.kr                
Business Name: Korean Society of Anesthesiologists (대한마취통증의학회)
Business Registration: 106-82-07194
Representative: Young-Tae Jeon (전영태)

Copyright © 2025 by Korean Society of Anesthesiologists.

Developed in M2PI

Close layer
prev next