Korean J Anesthesiol Search

CLOSE


Korean Journal of Anesthesiology 2004;47(3):442-445.
DOI: https://doi.org/10.4097/kjae.2004.47.3.442   
Uncommon Complications Related to Pulmonary Artery Catheterization (CCOmbo Swan-Ganz catheter): Two case report.
Jong Yeon Park, Suk Jin Lee, In Cheol Choi
1Department of Anesthesiology and Pain Medicine, College of Medicine, University of Ulsan, Seoul, Korea. icchoi@amc.seoul.kr
2Department of Anesthesiology and Pain Medicine, Gangneung Asan Medical Center, Gangneung, Korea.
Abstract
Case 1: A 30-year-old woman was scheduled for mitral valvuloplasty. A pulmonary artery (PA) catheter (Swan-Ganz CCOmbo model: 744HF75, Edwards(r), USA) was inserted via an introducer (8.5 Fr) placed in the right external jugular vein. The tip of the catheter could not be passed from the external jugular vein to the subclavian vein. When we attempted to advance the PA catheter again 30 minutes later, it could not be advanced or withdrawn. The PA catheter was removed with an introducer. The tip of the introducer had curled up into the intralummial space and the PA catheter had impacted in the narrowed lumen.
Case
2: A 48-year-old woman was scheduled for tricuspid annuloplasty. A PA catheter was inserted via an introducer placed in the right internal jugular vein without problem. After opening the right atrium and withdrawing the PA catheter, a vinyl strip was found to have torn away from the distal part of the PA catheter.
Key Words: complications; pulmonary artery catheter


ABOUT
ARTICLE CATEGORY

Browse all articles >

BROWSE ARTICLES
AUTHOR INFORMATION
Editorial Office
101-3503, Lotte Castle President, 109 Mapo-daero, Mapo-gu, Seoul 04146, Korea
Tel: +82-2-792-5128    Fax: +82-2-792-4089    E-mail: journal@anesthesia.or.kr                

Copyright © 2024 by Korean Society of Anesthesiologists.

Developed in M2PI

Close layer
prev next