Korean J Anesthesiol Search

CLOSE


Korean Journal of Anesthesiology 2005;48(3):232-234.
DOI: https://doi.org/10.4097/kjae.2005.48.3.232   
Novel Self-inflating Resuscitators Maintaining High FiO2 with High Minute Ventilation.
Jeong Rim Lee, Seung Yeon Yoo, Teo Jeon Shin, Jae Hyun Bahk, Byung Moon Ham, Yun Seok Jeon
Department of Anesthesiology and Pain Medicine, Seoul National University Hospital, Seoul, Korea. jeonyunseok@hanmail.net
Abstract
BACKGROUND
Self inflating resuscitators are widely used for the patients who need positive pressure ventilation during transport. During self inflating resuscitator ventilation, the FiO2 may decreases even with the use of the oxygen reservoir. Such phenomenon may increase the difficulty and risk of transport of the critically ill patients. Twin self inflating resuscitator was assembled by being modified from two conventional ones to achieve more stable FiO2.
METHODS
The twin self inflating resuscitator and one conventional adult self inflating resuscitator were tested on a test lung. In the twin resuscitator, two adult self inflating resuscitators were connected serially without distal unidirectional valve plate. The resuscitators were compressed at variable tidal volumes, respiratory rates and oxygen flow rates, guided by the monitor.
RESULTS
With conventional self inflating resuscitator ventilation, the FiO2 was maintained over 95% until minute ventilation of 7,500, 1,4000, 17,500 ml at respectively 5, 10, 15 L/min oxygen flow rate. With serial type, the FiO2 started to decrease with the minute ventilation over 12,500, 24,000, 28,000 ml at 5, 10, 15 L/min oxygen flow rate.
CONCLUSIONS
By simple connection of two self inflating resuscitators, the FiO2 during self inflating resuscitator ventilation could be maintained during almost two times of minute ventilation compared to conventional ones.
Key Words: intrahospital transport; self-inflating resuscitator


ABOUT
ARTICLE CATEGORY

Browse all articles >

BROWSE ARTICLES
AUTHOR INFORMATION
Editorial Office
101-3503, Lotte Castle President, 109 Mapo-daero, Mapo-gu, Seoul 04146, Korea
Tel: +82-2-792-5128    Fax: +82-2-792-4089    E-mail: journal@anesthesia.or.kr                

Copyright © 2024 by Korean Society of Anesthesiologists.

Developed in M2PI

Close layer
prev next