Korean J Anesthesiol Search

CLOSE


Korean Journal of Anesthesiology 2001;41(2):239-243.
DOI: https://doi.org/10.4097/kjae.2001.41.2.239   
Increased Peak Inspiratory Pressure Due to Intraluminal Bulging of the Inner Layer of the Reinforced Wire Tube during Anesthesia: A case report.
Hae Keum Kil, Bon Nyeo Koo, Mee Kyung Kim
Department of Anesthesiology, Yonsei University College of Medicine, Seoul, Korea.
Abstract
Excessive peak airway pressure during general endotracheal anesthesia may result from bronchospasm due to light anesthesia or surgical stimulation, bronchial intubation, tension pneumothorax, pulmonary edema, or mechanical obstruction of tube, whether from kinking, inspissated secretions, or overinflation of the cuff. Usually these problems are differentiated with auscultation and drug administration. However, mechanical problems associated with the endotracheal tube may be a cause of increased airway pressure. Reinforced, anode, or armored tubes consist of two coatings of latex or PVC that enclose spiral metal windings. Because of that, the inner layer may peel away, and intraluminally bulge due to nitrous oxide and cause airway obstruction during the course of an anesthetic process. We report a case of intraluminal bulging of the inner layer in a reinforced tube using fiberoptic bronchoscopy during anesthesia.
Key Words: obstruction; peak airway pressure; anode; armored reinforced wire tube


ABOUT
ARTICLE CATEGORY

Browse all articles >

BROWSE ARTICLES
AUTHOR INFORMATION
Editorial Office
101-3503, Lotte Castle President, 109 Mapo-daero, Mapo-gu, Seoul 04146, Korea
Tel: +82-2-792-5128    Fax: +82-2-792-4089    E-mail: journal@anesthesia.or.kr                

Copyright © 2024 by Korean Society of Anesthesiologists.

Developed in M2PI

Close layer
prev next