Korean J Anesthesiol Search

CLOSE


Korean Journal of Anesthesiology 2000;38(1):89-93.
DOI: https://doi.org/10.4097/kjae.2000.38.1.89   
The Sealing Volume and Pressure of the Bronchial Blocker Cuff of Univent(R) Tubes.
Rack Kyung Chung
Department of Anesthesiology, College of Medicine, Ewha Womans University, Korea.
Abstract
BACKGROUND
The manufacturer has recommended that 6-7 ml of air be used to seal the bronchial blocker cuff of Univent tubes high pressure type. However overinflation of the cuff might result in pressure damage to mucosa. This study was performed to evaluate the appropriate sealing volume and pressure of the bronchial blocker cuff.
METHODS
Univents were inserted in 26 patients. The bronchial blocker was connected to a pressure gauge and the cuff was inflated by 0.5 ml increment until the bronchus was sealed in supine and lateral positions under 20-25 mmHg airway pressure. The sealing volume and pressure were recorded.
RESULTS
In males, the sealing volume for the left bronchus (3.8 +/- 0.4 ml) was smaller than for the right bronchus (5.4 +/- 1.0 ml) in both positions significantly (p < 0.05). In females, the sealing volume for the left bronchus (2.8 +/- 0.4 ml in both positions) was smaller than the volume for the right bronchus (3.7 +/- 0.8 ml in supine position, 3.9 +/- 0.9 ml in lateral position) significantly (p < 0.05). The sealing volume for females was smaller than for males significantly (p < 0.05), for all sealing pressures, there were no significant differences statistically.
CONCLUSIONS
The bronchial blocker of Univent is a high pressure type and the sealing volume is higher for the right bronchus than for the left bronchus. Furthermore the sealing volume is less than the volume which is recommended by the manufacturer.
Key Words: Equipment: Univent; Measurement: sealing volume; sealing pressure


ABOUT
ARTICLE CATEGORY

Browse all articles >

BROWSE ARTICLES
AUTHOR INFORMATION
Editorial Office
101-3503, Lotte Castle President, 109 Mapo-daero, Mapo-gu, Seoul 04146, Korea
Tel: +82-2-792-5128    Fax: +82-2-792-4089    E-mail: journal@anesthesia.or.kr                

Copyright © 2024 by Korean Society of Anesthesiologists.

Developed in M2PI

Close layer
prev next