A practical guide to oropharyngeal leak pressure for supraglottic airway devices
Article information
Abstract
Supraglottic airway devices (SGAs) are routinely used for airway maintenance and are considered key rescue devices in difficult airway algorithms when tracheal intubation fails. The oropharyngeal leak pressure (OPLP) is the most practical bedside surrogate for assessing SGA sealing performance, as it approximates the upper limit of airway pressure during positive-pressure ventilation. In this narrative review, we summarize how OPLP should be interpreted and used in practice. We define OPLP, explain the clinical meaning of low and high OPLP for ventilation adequacy and gastric insufflation, present practical approaches to OPLP measurement, present sources of interstudy variability, and describe common pitfalls related to malpositioning and protocol differences. We summarize the comparative evidence for widely used SGAs, including the i-gel family, LMA Supreme, LMA ProSeal, Baska Mask, Air-Q, and Ambu AuraGain, and highlight scenarios in which device-level differences may become clinically important upon increased airway pressure (e.g., obesity, pneumoperitoneum, Trendelenburg positioning, or reduced lung compliance). We also review patient-, procedure-, and operator-related factors that modify the OPLP and propose a practical approach for troubleshooting borderline ventilation, including repositioning, adjusting cuff pressure, upsizing or switching devices, and escalating to tracheal intubation. Finally, we outline the priorities for standardized reporting and future directions for developing individualized airway strategies that integrate bedside seal metrics with patient-specific assessments.
Introduction
Supraglottic airway devices (SGAs) are routinely used for airway maintenance. They are strongly recommended as rescue devices when tracheal intubation fails; thus, reliable ventilation with an SGA is a recurring clinical requirement [1–3]. The oropharyngeal leak pressure (OPLP) is a practical bedside metric of SGA sealing performance [4–10]. Because an inadequate seal and SGA malposition can contribute to ventilation failure and gastric insufflation, objective OPLP assessment and position optimization have been emphasized in contemporary discussions on safe SGA use [8,11,12]. Current standardized methods have shown good reliability in adults and children, supporting use of the OPLP as a feasible clinical and research endpoint [13,14]. Moreover, comparative trials and meta-analyses have demonstrated device-level differences in OPLP that may matter when airway pressures increase due to physiological or surgical conditions [15]. In this narrative review, we summarize the clinical significance of the OPLP, practical approaches for measuring this metric, comparative OPLPs across commonly used SGAs, factors that modify the OPLP, implications for difficult airway management (DAM), and future directions for more individualized airway strategies [4,8,15,16].
Literature search and scope
We performed a targeted search of PubMed/MEDLINE, Embase, and the Cochrane Library (1990–2025) using combinations of keywords and subject headings related to SGAs and sealing performance, including supraglottic airway, laryngeal mask, oropharyngeal leak/seal pressure, airway sealing pressure, cuff pressure, and gastric insufflation. Reference lists of key guidelines, systematic reviews, and device comparison trials were screened manually to identify additional studies. As this was a narrative review, we did not apply a formal risk-of-bias tool. Instead, we prioritized randomized trials and meta-analyses for between-device comparisons, and included landmark physiological and imaging studies to clarify sealing mechanisms, malpositioning, or measurement limitations. Additionally, we focused on widely used SGAs for which robust comparative OPLP literature is available, and with clinical relevance to rescue and positive-pressure ventilation. These included the i-gel family (Intersurgical), LMA Supreme (Teleflex), LMA ProSeal (Teleflex), Baska Mask (Logial Health Products), Air-Q (AirLife), and AuraGain (Ambu), which have substantial comparative OPLP data and are commonly used in settings requiring rescue ventilation in difficult airway situations and positive-pressure ventilation.
Clinical importance of the OPLP
In the SGA literature, “OPLP” and “oropharyngeal seal pressure” are often used interchangeably. However, the measured value depends on the predefined endpoint used to identify the leakage. To maintain consistency across this review, we used “OPLP” as a general term and adopted an operational definition aligned with contemporary comparative research: the airway pressure obtained using a standardized manometric stability test (with or without audible leak confirmation) at a fixed fresh-gas flow and a predefined pressure cap. In this context, the OPLP reflects the pressure at which the circuit plateaus stably under standardized conditions, which best supports between-device comparisons. Other leak detection methods are described as protocol variations and may yield slightly different values.
Anatomical contact sites of SGAs
The sealing performance of an SGA reflects a multi-point apposition between the device cuff and the surrounding pharyngo-laryngeal structures, rather than a single position at the laryngeal inlet. Seal formation depends on how the device conforms to the tongue base, oropharynx, peri-epiglottic tissues, and hypopharyngeal structures around the laryngeal inlet, which explains why devices with similar fiber-optic appearances can yield different OPLPs when their contact patterns and pressure distributions differ [4]. Cadaveric endoscopy and dissection studies of the i-gel have shown that its noninflatable, anatomically shaped cuff conforms to the perilaryngeal anatomy, yielding consistent contact around the supraglottic structures and supporting a geometry-driven sealing mechanism [5]. In vivo magnetic resonance imaging also revealed device-specific anatomical profiles. In a randomized crossover comparison of the i-gel and the LMA Supreme, both occupied the hypopharyngeal space, but differed in their relationships with adjacent tissues [6]. The physiological data complemented these observations. Directly measured mucosal pressure and OPLP were related, and the ProSeal produced a better OPLP than did the classical LMA, without increasing the measured mucosal pressure, indicating that sealing quality depends on contact distribution and pressure magnitude [7]. Overall, the OPLP can be viewed as an integrated outcome of contact sites, distribution, and stability, which depend on the specific structures contacted, the extent of contact distribution, and the persistence of the contact pattern under changing conditions [4].
Definition and clinical significance of the OPLP
The OPLP is a functional measure of how well an SGA seals against the periglottic and pharyngeal tissues. Unlike peak inspiratory pressure, which is largely determined by lung and chest wall mechanics and ventilatory settings, the OPLP reflects the sealing interface between the device and pharyngolaryngeal tissues more directly. Clinically, it approximates the upper limit of airway pressure that can be applied without clinically relevant leakage occurring during positive-pressure ventilation. Because the OPLP is easy to measure and obtain at the bedside, it is widely used to confirm SGA performance after insertion, and acts as a primary outcome for device comparisons [8]. However, the OPLP recorded depends on the measurement endpoint used and the protocol details.
The OPLP is most relevant when implementation of positive-pressure ventilation is planned. If the peak inspiratory pressure approaches or exceeds the OPLP, leakage can result in delivery of a reduced tidal volume and can compromise ventilation [8]. The device design can expand the usable pressure range. For example, the ProSeal achieved a higher OPLP than did the classic LMA, while the directly measured mucosal pressures remained similar, suggesting that some 2nd-generation designs can improve functional sealing without necessarily having to increase the pharyngeal mucosal pressure [7].
The OPLP is also related to gastric insufflation. Gastric insufflation is multifactorial and depends not only on the seal, reflected by the OPLP, but also on device malpositioning, ventilatory settings, and the pressure gradient across the upper esophageal sphincter. Ultrasound-based trials have suggested that insufflation can occur even at commonly used airway pressures and may be influenced by the cuff pressure [11]. Moreover, a randomized crossover study further reported that the presence of a gastric tube inserted through the gastric channel of 2nd-generation SGAs was associated with a higher frequency of gastric insufflation during incremental pressure ventilation and OPLP measurement, indicating that clinical and measurement conditions can influence the observed events [17]. As insufflation can occur even at modest airway pressures, the OPLP should be interpreted as a ventilation-focused seal metric that is useful for planning and device comparison, but does not guarantee optimal positioning or aspiration protection [8,11].
Risks of low and high OPLP
Both low and high OPLP levels may have clinically relevant consequences. When the OPLP is low, leaks occur at modest airway pressures. As part of the delivered volume escapes around the device rather than inflating the lungs, the tidal volume can be reduced and controlled ventilation may not be possible when higher pressures are required. The finding that the ProSeal provides a higher OPLP than the classic LMA underscores the concept that a low OPLP limits ventilatory capability [7]. Practically, this can be understood by considering the “seal margin,” defined as the difference between OPLP and the inspiratory pressure required for effective ventilation (e.g., the peak inspiratory pressure). When this margin is small, clinically relevant leakage becomes more likely as the ventilatory demand increases (e.g., due to obesity, pneumoperitoneum, Trendelenburg positioning, or reduced lung compliance). A low OPLP is also relevant because leakage and rising airway pressure can promote insufflation of the esophagus and stomach, contributing to regurgitation and aspiration. In a study of children ventilated via the i-gel device without a gastric tube, gastric insufflation was assessed alongside the OPLP and peak inspiratory pressure, and demonstrated that insufflation can occur during routine ventilation when the OPLP is limited [18]. In adults, real-time ultrasound studies have similarly indicated that gastric insufflation varies with clinical conditions, including stepwise increases in airway pressure and OPLP [17].
Conversely, a high OPLP generally indicates a larger seal margin and can support controlled ventilation when the airway pressure rises; however, very high values, achieved through excessive cuff inflation, can increase local tissue compression. Studies in which mucosal pressure was measured directly have shown that sealing performance and mucosal pressure are related, but are not interchangeable, and vary with the device and anatomical location [7,9]. In routine practice, cuff pressures may exceed the commonly accepted upper limits unless they are actively monitored. This is relevant because prolonged pressure is a plausible contributor to mucosal injury and neuropraxia [19]. The findings of a systematic review of cranial nerve injuries associated with SGAs provided support for pressure- and position-related mechanisms and emphasized the need for appropriate SGA sizing, placement, and cuff-pressure management [20]. Moreover, high-sealing conditions may influence distal esophageal physiology. In a cadaver study, cuff overinflation and applied neck pressure reduced the esophageal leak pressure [21]. In a pediatric study investigating the relationship of the cuff pressure and esophageal barrier pressure, LMA use did not significantly change lower esophageal barrier pressure overall, but 2 of 20 children showed substantial decreases after insertion, which could increase reflux risk [22]. Accordingly, when a high OPLP is obtained, particularly if it seems disproportionate to the clinical setting, the intracuff pressure, device size, and insertion depth should be rechecked to ensure that a seal is achieved without causing unnecessary pharyngeal compression.
How to measure the OPLP
Measurement methods
After SGA insertion, the OPLP is typically assessed using a standardized bedside procedure, most commonly by means of manometric stability testing in comparative trials and meta-analyses. Briefly, under a fixed fresh-gas flow rate and with a predefined pressure cap for safety, the adjustable pressure-limiting (APL) valve is gradually closed to allow the airway pressure to rise until the circuit pressure reaches a stable plateau (manometric equilibrium) or until a clearly detectable oral leak occurs. If the latter occurs before a stable plateau is reached, the pressure at the first detectable leak is recorded and represents the lower bound of the device’s sealing performance under that protocol. This distinction is important because plateau-based values tend to be slightly higher than leak detection endpoints, even under similar conditions [8]. Because the protocol details materially influence the measured OPLP, we have provided a pragmatic reference protocol and a standardized minimum set of reporting items in Table 1.
In anesthetized, paralyzed adults, Keller et al. [13] compared four methods for measuring OPLP with the LMA: audible mouth leak, end-tidal CO2 detection at the mouth, manometric stability, and neck auscultation, and found excellent inter-observer reliability across methods. Importantly, the manometric stability approach yielded slightly higher sealing pressures and more reliable results for comparison. Similar feasibility and reliability have been demonstrated for manometric stability evaluation in pediatric patients [14]. In this pediatric study, the measured values showed no clear systematic differences across the four methods, supporting the flexibility in the method used when the protocol was otherwise standardized. Therefore, manometric stability assessment is generally preferred and is most frequently reported in contemporary comparative studies, whereas the first-leak endpoints are more protocol-dependent and can yield lower values.
In contemporary trials and meta-analyses comparing SGAs, the OPLP has most often been determined by evaluating manometric stability, sometimes combined with detection of an audible mouth leak, which facilitates device-to-device comparisons across studies [8,15]. Nevertheless, protocols vary across studies, even within manometric procedures. The differences include the fresh-gas flow rate, maximum pressure allowed during assessment, and whether the cuff pressure was standardized or adjusted before measurement. These details can influence the recorded OPLP and contribute to interstudy variability.
In addition to OPLP assessment, imaging-based and endoscopic approaches have been explored as adjuncts for evaluating seal adequacy and device position. Ultrasound can be used to assess device seating in multiple planes and the results have been reported to agree with those of pressure-based assessments for identifying suboptimal SGA placement. Fiber-optic assessment is commonly used as a reference when needed [23,24]. These approaches provide different types of information. Pressure-based assessment yields a quantitative pressure threshold, whereas ultrasonography and bronchoscopy provide anatomical confirmation of the device position and may help to explain unexpectedly low OPLPs when insertion was uncomplicated. Procedure-specific physiological conditions can also alter the measured OPLP. During laparoscopic surgery in the lateral position, the OPLP with ProSeal increased after pneumoperitoneum, indicating that the OPLP can change with the evolution of intraoperative conditions [25].
Because OPLP assessment transiently increases airway pressure by closing the APL valve at a fixed fresh-gas flow rate, measurements should be performed briefly, with a predefined pressure cap, to avoid excessive insufflation or pressure-related complications. Clinical protocols commonly limit the maximum circuit pressure to 40 cmH2O and terminate the test once a clinically relevant leak is detected or a plateau is reached [26,27]. When gastric insufflation is suspected, ventilation is unstable, or the aspiration risk is high, clinicians should cautiously interpret the need for OPLP testing and prioritize optimizing device positioning, appropriate size selection, and intracuff pressure adjustment, rather than pursuing repeated high-pressure measurements.
Limitations and interstudy variability
Although the OPLP is a conceptually simple outcome, reported values have varied across studies because measurement procedures and clinical conditions have not been uniform. Studies differ in OPLP measurement conditions, even when the same four detection methods described above are used. The endpoint may be the first clinically detectable leak, the airway pressure at which it reaches a stable plateau, or a combination of these criteria. This can shift the measured value by several cmH2O in otherwise similar settings [13,14]. Studies have also varied in protocol parameters, such as the fresh-gas flow rate, the use of a maximum pressure cap for safety, and whether the assessment is performed during apnea or with ongoing spontaneous respiratory effort [8]. These differences complicate the comparisons of absolute OPLP values and can contribute to heterogeneity in meta-analyses.
The OPLP is measured under the assumption that the device is appropriately seated in the hypopharynx. However, blind SGA placement is often imperfect in routine practice. Therefore, malposition can contribute to between-study variability, particularly when fiber-optic confirmation is not routinely performed or when studies use different criteria to define acceptable positioning [8,12]. Anesthetic techniques and airway conditions can also influence the seal of the SGA. Neuromuscular blockade, depth of anesthesia, and pharyngeal muscle tone can affect seal stability during assessment and hence the reproducibility of repeated measurements [12]. In a meta-analysis of randomized trials comparing the ProSeal and i-gel devices, statistically significant heterogeneity was reported for OPLP. Differences in patient characteristics, anesthetic techniques, and measurement procedures were considered likely sources of variability [28]. Overall, the OPLP remains useful for within-study comparisons when the procedures are standardized. Cross-study interpretation benefits from the explicit reporting of measurement parameters, positioning checks, and anesthetic conditions.
Comparing OPLP between various SGAs
To compare SGA devices, Table 2 summarizes the representative OPLP values reported by device, together with the key study conditions, including population, ventilation mode, and use of neuromuscular blockade.
i-gel and i-gel Plus
The i-gel is a 2nd-generation single-use SGA with a noninflatable cuff that has been designed to conform to the perilaryngeal anatomy. In adults, the OPLP with the i-gel is generally adequate for routine positive-pressure ventilation, although it was lower than that achieved with the ProSeal in a meta-analysis (mean difference 3.37 cmH2O across 10 RCTs [29]; weighted mean difference about 1.5 cmH2O across 33 RCTs [15]). In these analyses, the insertion characteristics were broadly comparable, with the i-gel often showing easier insertion, less blood staining on the device, a shorter insertion time, and fewer postoperative sore throats [15,29]. When higher airway pressures are anticipated, the ProSeal may provide a larger sealing margin, whereas in routine cases, the i-gel offers practical advantages [15,29].
In adults with a normal airway anatomy, a 3-arm randomized trial reported no significant difference in the mean OPLP among the Air-Q, i-gel, and Supreme systems, although the OPLP for the i-gel was numerically lower in that cohort [30]. Device performance can be condition-dependent. Under the 60° head-rotation condition, the ProSeal maintained a higher OPLP than did the i-gel, regardless of the similar OPLPs for these devices under the neutral position condition [31]. In spontaneously breathing children, neck flexion increased, while extension decreased the OPLP for both the i-gel and Supreme, and the i-gel showed an OPLP that was higher than that of Supreme across positions [32].
The i-gel Plus is a newer modification of the i-gel; the design changes include a wider gastric drainage port, a longer cuff tip, ramps within the airway channel to facilitate intubation, and an additional port for oxygen delivery. In a large, prospective, international, multicenter study in adults, Werner et al. [33] reported high overall insertion success (98.6%) and a mean OPLP of 32 ± 7 cmH2O, with low rates of major complications with i-gel Plus. Operator inexperience was the primary independent predictor of first-attempt insertion failure. Presently, the evidence for the i-gel Plus is largely observational, and direct comparisons of its OPLP with those of established SGAs remain limited [33].
Clinically, the i-gel is a practical 2nd-generation SGA that is useful when rapid placement and consistent geometry-based seating are prioritized in routine and rescue cases. Because its noninflatable seal is more sensitive to positional stress in some settings, its OPLP should be interpreted in the context of head or neck position and intraoperative changes in airway-pressure demand.
LMA Supreme
The LMA Supreme is a single-use, 2nd-generation laryngeal mask with a preformed curved airway tube and an integrated gastric drain. Reported sealing pressures are commonly around 25 cmH2O in adults, but its comparative performance depends on the ventilation mode and study protocol [34–37]. In a randomized trial conducted in spontaneously breathing adults, the OPLP was higher with the Supreme than with the i-gel (mean 25.6 vs. 20.7 cmH2O) [34]. In elective surgery under controlled ventilation, a randomized trial found that the ProSeal achieved the highest OPLP, while the OPLPs of the Supreme and i-gel were more similar (mean OPLP: 32.6 cmH2O for ProSeal, 29.8 cmH2O for Supreme, and 26.7 cmH2O for i-gel) [35]. In laparoscopic cholecystectomy, the Supreme yielded a lower OPLP than that of the ProSeal (26.8 ± 4.1 vs. 30.7 ± 6.2 cmH2O), while first-attempt insertion success favored the Supreme [36]. Other randomized comparisons that included fiber-optic assessment reported similar leak pressures for the Supreme cuff and i-gel when the Supreme cuff pressure was standardized at 60 cmH2O [37]. In pediatric patients, the findings have been mixed. A randomized equivalence trial showed lower leak pressures with the Supreme than with the i-gel (median 17 vs. 20 cmH2O) [38]. Another pediatric RCT reported similar OPLPs between the Supreme and ProSeal (16.4 ± 1.7 vs. 17.2 ± 2.3 cmH2O) [39]. A pediatric meta-analysis concluded that the Supreme provides an OPLP comparable to that of the ProSeal and i-gel, with wide confidence intervals across studies [40].
Clinically, the Supreme offers predictable handling with integrated gastric access and adequate sealing for standard-controlled ventilation under standardized cuff-pressure conditions. The measured OPLP is more protocol-dependent when cuff-pressure management differs; therefore, explicit reporting and consistent cuff-pressure control are particularly important in comparative studies.
LMA ProSeal
The LMA ProSeal was developed to improve the airway seal during positive-pressure ventilation, while adding a gastric drainage channel. It remains a reference device for OPLP comparisons across 2nd-generation SGAs. Evidence syntheses have consistently shown that the OPLP of the ProSeal is higher than that achieved with the i-gel in adults, with a modest but reproducible pooled advantage [15,29]. Earlier RCTs comparing the ProSeal with the LMA Classic also demonstrated a higher OPLP with the ProSeal, supporting its ability to withstand higher airway pressures before clinically relevant leakage occurs [41,42]. Head and neck conditions can further determine the device performance. In a randomized trial of head rotation, the ProSeal maintained a higher OPLP than did the i-gel at 60° rotation, whereas the OPLP values in a neutral position were comparable, suggesting that the ProSeal can be more resistant to seal degradation during extreme positioning [31]. During laparoscopy, multiple randomized studies that compared the ProSeal with the LMA Supreme reported higher OPLPs with the ProSeal under standardized cuff-pressure conditions, including during laparoscopic cholecystectomy [36,43]. The OPLP can also change with intraoperative physiology. In laparoscopic gynecological surgery, pneumoperitoneum and Trendelenburg positioning alter the measured OPLP for both the ProSeal and i-gel, reinforcing that the ProSeal sealing performance should be interpreted in the context of operative conditions rather than as a fixed device property [44]. In pediatric practice, the ProSeal performance is size-dependent, because smaller sizes (1.5–2.5) lack a dorsal cuff. In a randomized pediatric trial, ProSeal provided a lower OPLP than did the AuraGain, which may reflect differences in the design of smaller airways [45].
Clinically, the ProSeal remains a standard high-seal device for use when a larger seal margin is anticipated. It is most clinically relevant when airway pressure is expected to increase and gastric access is desired without sacrificing sealing performance.
Baska Mask
The Baska Mask is a newer SGA with a noninflatable, self-sealing membranous cuff that is designed to augment the seal as the airway pressure rises. In a study on laparoscopic cholecystectomy, the Baska Mask produced a higher mean OPLP than did the i-gel mask under adequate ventilation during pneumoperitoneum [46]. Another randomized study also reported higher OPLP with the Baska Mask than with the i-gel, while the overall clinical performance remained acceptable [47]. In elective surgery, the Baska Mask achieved an OPLP comparable to or higher than that of the i-gel, although their insertion characteristics and associated postoperative sore throat differed [48].
Comparisons with 2nd-generation cuffed LMAs have revealed similar patterns. During controlled ventilation, the Baska Mask achieved a higher OPLP than did the LMA Supreme [49]. Randomized studies have further reported a higher OPLP with the Baska Mask than with the ProSeal under both intermittent positive-pressure ventilation and controlled ventilation [50,51]. In adult patients with obesity, a 4-arm randomized study found that, among i-gel, AuraGain, Baska Mask, and LMA Protector, the Baska Mask yielded the highest mean OPLP [52]. A meta-analysis that compared the Baska Mask with the i-gel concluded that the pooled OPLP of the Baska Mask was at least comparable and often higher than that of the i-gel, although it possibly increased the postoperative sore throat incidence [53].
Clinically, the Baska Mask is a high-seal option that may perform well as the airway-pressure demand increases, as demonstrated in comparative studies. As insertion characteristics and postoperative throat symptoms vary across reports, its selection should balance the priority of good seal with tolerability and operator familiarity.
Air-Q
The Air-Q family of SGAs, including the intubating laryngeal airway and the self-pressurizing Air-Q (Air-Q SP), is designed to function as both a primary supraglottic airway and a conduit for tracheal intubation; therefore, the sealing performance of these devices must be interpreted alongside intended use. In a study of adults with normal airway anatomy, the mean OPLP did not differ significantly among the Air-Q, i-gel, and LMA Supreme groups, with values in the mid-20 cmH2O range [30]. In an earlier randomized trial, the OPLP was similar between the Air-Q and ProSeal in adults undergoing elective outpatient surgery [54], whereas a more recent randomized study reported a higher mean OPLP with the ProSeal under effective ventilation conditions with both devices [55]. Pediatric data have suggested a smaller sealing margin with Air-Q SP than with some 2nd-generation SGAs. The Air-Q SP had a lower OPLP than that of the Supreme in children aged 1–7 years [56]. In another study, the i-gel yielded a higher OPLP and was associated with less gastric insufflation than the Air-Q SP, whereas the Air-Q SP offered a better fiber-optic view [57].
Clinically, the Air-Q is most useful when an intubation-conduit strategy is a key objective, while still providing acceptable sealing in many routine settings. In pediatrics, its sealing performance should be interpreted alongside its facilitation of fiber-optic viewing and intubation rather than by the OPLP alone.
Ambu AuraGain
The Ambu AuraGain is a 2nd-generation single-use SGA that features an anatomically curved airway tube, an inflatable cuff, a gastric channel, and a conduit that facilitates intubation. In adults undergoing ambulatory surgery, Wong et al. [58] reported a higher OPLP with the AuraGain than with the Supreme, although placement of the former took longer (13 ± 4 vs. 11 ± 3 s). In laparoscopic cholecystectomy, the AuraGain and the i-gel generated similar OPLP values and provided controlled ventilation during pneumoperitoneum, with comparable perioperative complications [59]. In a study involving laparoscopic inguinal hernia repair, Anupma et al. [60] observed that the AuraGain maintained a higher OPLP than did the i-gel at 30 min after pneumoperitoneum and positioning, whereas the i-gel could be inserted faster. In laparoscopic gynecological surgery, the OPLP with the AuraGain increased after pneumoperitoneum and Trendelenburg positioning, emphasizing that intraoperative conditions can modify sealing performance [61]. In adults with obesity, Raman et al. [52] reported that the AuraGain had a high OPLP, comparable to that of the i-gel and LMA Protector, although the Baska Mask achieved the highest OPLP. Pediatric data also support the sealing performance of the AuraGain. In a randomized trial in children under controlled ventilation conditions, the AuraGain achieved a higher OPLP than that of the ProSeal and facilitated gastric tube passage [45]. Randomized pediatric comparisons of the AuraGain with the LMA Supreme and i-gel have shown that it produces an effective seal and is associated with high ventilation success [62,63]. A recent non-inferiority study suggested that the AuraGain maintains the OPLP with the patient in the prone position when intracuff pressure is rechecked after repositioning [64].
Clinically, the AuraGain is a versatile 2nd-generation SGA when both a solid seal and intubation through the device are desired, with consistent evidence for a robust OPLP and gastric access. As this SGA is cuffed, standard intracuff pressure management and reassessment after patient repositioning are central to maintaining reproducible performance.
Taken together, beyond OPLP values, device selection should also consider device-specific strengths and limitations, as well as the clinical context in which each device performs best. These practical considerations are summarized in Table 3.
Factors influencing the OPLP
This section outlines the major determinants of the OPLP in clinical practice. A concise summary of these factors and their typical directional effects is presented in Table 4.
Sex
Sex effects on the OPLP in adults often reflect pharyngeal size and device size, rather than sex per se. In an LMA Classic crossover study, men had a higher OPLP with size 5 than with size 4, whereas women had similar OPLP with sizes 5 and 4, indicating that undersizing is more likely to reduce the seal in men [65]. For the LMA ProSeal, randomized crossover data support sex-based size selection to optimize the OPLP under standardized conditions [66]. In an Asian population, ProSeal sizing according to sex was similarly associated with clinically meaningful differences in the OPLP [67]. These findings suggest that appropriate SGA size selection is key to optimizing the OPLP across sexes.
Obesity
Obesity can increase airway-pressure requirements, raising concern that the SGA seal may be insufficient for controlled ventilation. Available clinical data suggest that, with modern 2nd-generation devices and appropriate SGA sizing, the OPLP can often be preserved in patients with mild-to-moderate obesity. In a prospective study comparing adults with and without obesity, the i-gel achieved a similar OPLP in both groups and enabled successful ventilation in all enrolled patients with obesity, although patients with very severe obesity and high aspiration risk were excluded from the study [68]. In a randomized crossover trial in adults with mild-to-moderate obesity, the i-gel produced a higher OPLP than that of the LMA Unique, indicating that newer device designs can provide a larger sealing margin than that of the older cuffed devices in this population [69]. Evidence specific to the ProSeal in adults with obesity also supports its use in positive-pressure ventilation in this population. In a randomized trial that compared the ProSeal with the LMA Classic in patients with a body mass index > 30 kg/m2 and that used progressive testing to define excessive leakage under standardized cuff pressure, the ProSeal was shown to sustain ventilation requirements in patients with obesity [70]. Physiological investigations further showed that the OPLP remained in the mid-20–30-cmH2O range with the ProSeal in patients with obesity under pressure-controlled ventilation, with oxygenation benefits observed when positive end-expiratory pressure (PEEP) was added [71]. A Cochrane review concluded that SGAs, particularly 2nd-generation devices, can be effective alternatives to tracheal intubation in selected patients with obesity, although the quality of evidence and generalizability of the risk of aspiration remain important considerations [72].
Pediatrics
The OPLP in children should be interpreted in the context of pediatric upper-airway anatomy, device sizing, and the typically lower inspiratory pressures used for ventilation. In pediatric randomized and comparative studies, 2nd-generation SGAs generally provided OPLPs that were sufficient for routine pediatric positive-pressure ventilation [45,62,63]. A network meta-analysis found that the ProSeal had a high OPLP with a low risk of device failure, while the i-gel also had a high OPLP but a higher overall device-failure rate; first-attempt insertion failure was also more frequent in children weighing < 10 kg [73]. A pediatric systematic review and meta-analysis reported a higher OPLP with the i-gel than with several LMAs, on average [74]. Individual trials have reported mixed findings. One study reported that the i-gel and ProSeal can yield similar OPLPs under controlled ventilation in certain age/size ranges [75], whereas another pediatric RCT reported a higher OPLP with the i-gel than with the ProSeal under standardized measurement conditions [76].
The OPLP can increase over the first 10 min after insertion in children, consistent with progressive seating and adaptation to the surrounding tissues [62]. Design differences for smaller sizes may also be important. The AuraGain achieved a higher OPLP than did the ProSeal in a pediatric RCT: smaller ProSeal sizes lack a dorsal cuff, which may limit sealing in smaller children [45]. Pediatric OPLP is also sensitive to head and neck positioning, and flexion and extension change the OPLP; thus, measurements should be obtained in a neutral position when possible [32].
Patient position
Patient position can significantly alter the OPLP, particularly when patients are turned after SGA placement. In children, changing from the supine to the lateral decubitus position reduced the OPLP for both the i-gel and ProSeal by roughly one-quarter and increased fractional volume loss, indicating a narrower margin for controlled ventilation with the patient in the lateral position [77]. Similar effects were observed in adults. During lateral laparoscopic surgery, the OPLP was consistently higher with the patient in the supine position than with the patient in the lateral position for both the ProSeal and Supreme, and pneumoperitoneum did not further decrease the OPLP beyond the positional effect [78]. Prone positioning is feasible when high-sealing devices are used. In adults in the prone position, the ProSeal achieved a higher OPLP than did the i-gel and provided a larger seal margin for positive-pressure ventilation [79]. In children in the prone position, the AuraGain maintained a non-inferior OPLP as compared with that in the supine position [64]. A clinical series also supported the use of the Supreme for providing clinically acceptable ventilation in adults in the prone position [80].
Head position
Head and neck alignment can change the pharyngeal shape around an SGA and can alter the OPLP, even when the device appears to be well-seated. In adults who received neuromuscular blockade, both the LMA Protector and the ProSeal yielded higher OPLPs in flexion and slightly higher values with a 30° head rotation, whereas extension consistently reduced the OPLP [81]. With a more extreme head rotation, seal deterioration became more apparent. In adults, the OPLP decreased with head rotation from 30° onward, although the ProSeal maintained a higher OPLP than did the i-gel at 60° [31]. Pediatric data showed the same pattern: flexion increased the OPLP, while extension decreased it for both the i-gel and the Supreme, although extreme flexion still impaired ventilation, despite the higher OPLP [32].
Laryngopharyngeal abnormalities
Structural distortion of the laryngopharynx can reduce the OPLP by preventing uniform device apposition, even when insertion appears to have been successful. In adults with post-burn neck contractures, the i-gel provided higher OPLPs and overall success than did the LMA Classic, suggesting that the anatomic seal design may better tolerate altered upper-airway geometry in selected patients [82]. In patients with cervical immobilization by means of a rigid collar, the Supreme maintains a clinically useful OPLP, supporting its use when neck movement is restricted [83]. Large audits and incident analyses have emphasized that, when the upper airway is grossly abnormal, SGAs may fail to seal or ventilate, despite correct placement attempts; therefore, early transition to an alternative airway strategy is essential [1].
Use of muscle relaxants
Neuromuscular blockade is often used to facilitate SGA insertion and control ventilation; however, its effect on the OPLP is inconsistent. In a randomized trial with the ProSeal, rocuronium increased the OPLP, suggesting that a reduced upper-airway tone can improve the seal in some patients [84]. In contrast, Goldmann et al. [85] found no significant change in the mean ProSeal leak pressure after establishing complete paralysis, although a minority showed a clinically relevant decrease in the OPLP, which supports the need to recheck the OPLP after neuromuscular blockade administration. In older adults using the i-gel, the OPLP was similar with and without rocuronium administration [86]. Meta-analysis data also indicated that neuromuscular blockade has heterogeneous effects, rather than reliably increasing the OPLP [15]. In practice, neuromuscular blockers should be chosen according to surgical conditions and should be used to suppress coughing or straining, rather than to increase OPLP.
Operator experience and technique
Operator-related factors influence the OPLP mainly by achieving optimal device seating rather than by experience alone. In a randomized trial, laryngoscope-guided LMA placement resulted in a significantly higher OPLP than that achieved with blind insertion, supporting the idea that visual guidance can help to correct subtle malpositioning when leaks are encountered [87]. A recent network meta-analysis also noted that the heterogeneity in SGA performance across trials is partly explained by operator proficiency and device familiarity, reinforcing the value of standardized training and consistent insertion protocols [88]. If the OPLP is unexpectedly low, the device should be repositioned and the OPLP rechecked before switching the device.
Use of lubricants or adjuncts
Appropriate lubrication supports smooth placement; however, the seal depends mainly on the final device seating, rather than on the lubricant itself. Early fit may be enhanced by device preparation. In a randomized, double-blind study, prewarming the Streamlined Liner of the Pharynx Airway improved alignment with laryngeal structures and increased the early OPLP during the first minute after insertion [89]. For cuffed LMAs, a randomized study found no significant difference in leak grade assessed at an airway pressure of 20 cmH2O, between insertion with a partially inflated cuff and insertion with a fully deflated cuff [90].
Impact of pneumoperitoneum on OPLP
During pneumoperitoneum, the peak inspiratory pressure increases. Therefore, the key criterion is that the OPLP exceeds the expected peak inspiratory pressure with a practical safety margin. In laparoscopic cholecystectomy in adults, the i-gel and the LMA Supreme showed similar OPLPs, which both increased after pneumoperitoneum as compared with immediate post-insertion values, suggesting that seating improves once ventilation and pneumoperitoneum are established [91]. In pediatric laparoscopy, three 2nd-generation SGAs (i-gel, LMA Supreme, and LMA ProSeal) maintained clinically useful leak pressures under pneumoperitoneum, indicating that when the OPLP is checked and remains above the set peak inspiratory pressure, controlled ventilation is supported [92]. This relationship is most relevant for the Trendelenburg position or under higher insufflation pressures.
Intraoperative or time-course changes
The OPLP can change after device placement; therefore, timing of OPLP assessment matters. In anesthetized, paralyzed adults, repeated leak pressure measurements over 30 min showed a stable OPLP, and nitrous oxide did not measurably affect the OPLP [93]. In pediatric cases, the OPLP increased by approximately 5 cmH2O from early measurement to the end of surgery, consistent with warming and improved seating of the device [94]. Therefore, measuring the OPLP after initiating ventilation and rechecking it if the early values are borderline are recommended.
Role of OPLP in DAM
SGAs as rescue devices in failed intubation
In DAM, the immediate priority is oxygenation and confirmation of effective ventilation. The OPLP is not a first-line decision metric during the crisis phase. Only once rescue ventilation has been established and the physiology has stabilized, does the OPLP become a practical bedside indicator, helping clinicians to judge whether positive-pressure ventilation can be sustained through an SGA and whether escalation should be pursued, such as SGA-guided intubation, awakening, or an alternative airway strategy. Currently, DAM algorithms position 2nd-generation SGAs as the key rescue step after failed tracheal intubation, aiming to re-establish oxygenation quickly and create time for a definitive strategy. Formal OPLP measurements can be performed once ventilation has been stabilized. In both the 2022 American Society of Anesthesiologists and the 2015 Difficult Airway Society guidelines, SGA insertion is recommended when further laryngoscopy attempts are unlikely to succeed and mask ventilation is inadequate [2,3]. Clinically, the seal matters because effective rescue ventilation requires the OPLP to exceed the inspiratory pressure required for the chest to rise, and a higher OPLP reduces leakage during positive-pressure ventilation. A large multicenter analysis showed that the use of SGAs as a rescue technique after failed direct laryngoscopy was frequently successful and comparable to other rescue options in practice [95]. After rescue ventilation is achieved, the OPLP should be confirmed, followed by a clear next step, such as SGA-guided intubation or patient awakening.
OPLP as a predictor of ventilation adequacy in DAM
After rescue ventilation is established, the OPLP provides a rapid and accurate estimate of the effectiveness of positive-pressure ventilation. If the peak inspiratory pressure required to ventilate approaches the measured OPLP, leakage around the device increases, the delivered tidal volume decreases, and gastric insufflation becomes more likely [94,96]. Conversely, when the OPLP exceeds the required peak airway pressure with a clear safety margin, ventilation is usually stable and gastric distension is uncommon, supporting its use as a temporary bridge while deciding the next step, such as waking the patient or intubating through the SGA [94]. Meta-analytical data have indicated that the ProSeal tends to achieve a higher OPLP than does the i-gel, which may matter in borderline physiological cases, such as patients with obesity or reduced respiratory system compliance [15]. The intensive care unit airway literature also cautions that, when pressures above roughly 30 cmH2O are needed, many SGAs cannot maintain a seal; therefore, a low OPLP should prompt early escalation to a definitive airway plan, rather than using persistent high-pressure ventilation through the SGA [96].
Practical implications of high OPLP for controlled ventilation in DAM
A higher OPLP is valuable in DAM because it can provide a more reliable bridge for controlled ventilation. The team can plan the definitive next step, particularly when airway pressures are elevated. Controlled ventilation often requires a higher peak pressure and PEEP in patients with reduced respiratory compliance or increased airway resistance. 2nd-generation SGAs that achieve a higher OPLP are therefore more likely to maintain stable tidal volumes without suffering progressive leakage or gastric insufflation, whereas lower-seal devices may become nonfunctional as ventilatory demands increase [97]. Evidence from laparoscopic surgeries supports this practical rule: SGA ventilation is generally acceptable when the OPLP remains above the peak inspiratory pressure. This emphasizes that the OPLP is a bedside predictor of whether controlled ventilation can be sustained after exposure to physiological stressors, such as pneumoperitoneum or steep positioning [29]. Device comparison studies have further indicated that even a modest OPLP advantage can be decisive in borderline scenarios [15]. Clinically, the OPLP should be measured soon after rescue SGA placement and should be reassessed after major changes in position or mechanics to determine whether ongoing controlled ventilation is reasonable or whether escalation is required [8].
Limitations and risks of SGAs with a low OPLP in DAM scenarios
In DAM, an SGA with a low OPLP indicates an incomplete seal and can become a liability. A low OPLP commonly leads to inadequate tidal-volume delivery when higher airway pressures are needed and promotes gastric insufflation with subsequent regurgitation and aspiration risks because the airway is not secured [96]. When oxygenation remains marginal after SGA placement, repeated attempts to ventilate through a leaking device can be time-consuming and may lead to a false sense of progress. The Fourth National Audit Project of the Royal College of Anaesthetists and the Difficult Airway Society highlighted the poor outcomes, including ongoing hypoxemia and regurgitation, resulting from persistence with an ineffective SGA, which supports early transition to a higher-sealing 2nd-generation device or an emergency surgical airway if ventilation is not rapidly restored [1]. Cognitive airway tools reinforce this principle: the Vortex approach emphasizes the rapid assessment of whether a face mask, SGA, or tracheal tube ventilation is working. The related commentary stresses that a failed SGA insertion attempt should trigger timely escalation to front-of-neck access, rather than prolonged troubleshooting [98,99]. Repeated high-pressure leakage testing in unstable situations is unlikely to help, or may worsen gastric insufflation. Clinicians should treat a low OPLP as an objective warning sign, confirm it with a leakage test and capnography, attempt to optimize placement once, and escalate without delay if oxygenation does not improve.
Future perspectives
Research gaps
Although the OPLP is widely used to judge the quality of an SGA seal, the evidence base for how the measured OPLP should guide real-time clinical decisions remains limited, particularly in patients with elevated airway-pressure demands and in settings where conditions change dynamically during anesthesia [2]. A major gap in knowledge is the lack of standardized OPLP measurements and reporting, including gas flow, APL settings, cuff-pressure management, head or neck position, and timing after insertion, which complicates cross-study comparisons and meta-analytic inferences [1]. Current comparative evidence is heterogeneous. Even when meta-analyses show device-level differences in leak pressure, the clinical meaning of a difference of a few cmH2O remains uncertain in the absence of outcome-linked thresholds, such as the adequacy of ventilation, hypercapnia, gastric insufflation, or aspiration risk [29,88]. Future studies should prioritize clinically relevant investigations and registries that pair the OPLP with airway-pressure demands and clinically important endpoints, and should test actionable thresholds for repositioning, adjusting the depth of anesthesia, upsizing, switching devices, adding adjuncts, or converting to tracheal intubation. To improve cross-study comparability, investigators should document and report the OPLP measurement conditions by using the standardized items listed in Table 1.
Toward personalized airway strategies
Personalized airway strategies should complement standard airway management by matching device choice and placement techniques with individual anatomical and ventilatory demands [2,3]. Point-of-care airway ultrasonography can provide patient-specific information on tongue thickness, hyomental distance, and upper-airway pathology, which will help clinicians anticipate SGA fit, select an appropriate SGA type and size, and choose tracheal intubation when needed [100]. Emerging computer vision and 3-dimensional facial scanning may further quantify facial morphology and predict difficult mask ventilation or suboptimal SGA fit, thereby enabling data-driven device sizing and positioning [101,102]. Future decision support that combines these features with bedside measurements, such as OPLP, could guide safer, individualized positive-pressure ventilation strategies and prompt early adjustments when the seal margin is low.
Conclusions
The OPLP is a practical, ventilation-focused metric that reflects how well an SGA conforms to perilaryngeal structures and tolerates changes in intraoperative conditions. Across devices, a higher OPLP generally provides a wider margin for controlled ventilation, while a low OPLP indicates malpositioning, inadequate seating, or a mismatch between the device and the patient’s airway anatomy, which can compromise tidal-volume delivery and promote gastric insufflation. Because the OPLP varies with the measurement method used, the patient’s head and body position, muscle tone, and the time after insertion, clinicians should use standardized bedside testing and should reassess the OPLP after major physiological or positional changes of the patient. Future studies should link OPLP thresholds to clinically important outcomes and should focus on personalized, data-guided airway strategies.
Notes
Acknowledgements
The authors used ChatGPT as an AI-assisted tool in the writing of this manuscript. All content was reviewed and finalized by the authors, who take responsibility for the manuscript in its entirety.
Funding
None.
Conflicts of Interest
No potential conflict of interest relevant to this article was reported.
Data Availability
Data sharing not applicable to this article as no datasets were generated or analyzed during the current study.
Author Contributions
Kota Saito (Conceptualization; Data curation; Formal analysis; Project administration; Validation; Visualization; Writing – original draft)
Soichi Tanaka (Conceptualization; Data curation; Formal analysis; Project administration; Validation; Visualization; Writing – original draft)
Tomohiro Chaki (Conceptualization; Data curation; Formal analysis; Investigation; Methodology; Supervision; Validation; Writing – review & editing)
Michiaki Yamakage (Conceptualization; Formal analysis; Methodology; Project administration; Supervision; Writing – review & editing)