Dear Editor,
We read with great interest the study conducted by Mostafa et al. [
1], which examined the diagnostic value of the ultrasound-measured skin-to-hyoid bone distance (SHD) for predicting difficult mask ventilation (DMV) in patients with obesity. The prospective design, relatively large sample size, and use of k-fold cross-validation represent valuable contributions to the literature. However, we would like to raise three points that, in our view, were not addressed by the authors as a limitation.
First, the authors excluded patients with suspected difficult laryngoscopy who had “suspected DMV.” While this approach is understandable on clinical safety grounds, the a priori exclusion of the highest-risk cases may have influenced both the observed incidence of DMV (6.7%) and the generalizability of the SHD cutoff value. Indeed, a recent meta-analysis reported a DMV incidence of approximately 14.4% in obese populations, which is considerably higher than that observed in the present cohort [
2]. The exclusion of the highest-risk patients raises the question of how well the 1.9-cm SHD threshold and its negative predictive value would perform in an unselected, “real-world” obese population. We believe that this point merits inclusion as a limitation, given its relevance to the clinical translatability of the findings.
All ultrasound measurements in this study were performed by a single experienced anesthetist, and the reported intraclass correlation coefficient (0.99) reflected only the intra-rater consistency of three repeated measurements by the same operator. Inter-rater reliability across operators with varying levels of experience was not assessed. The airway ultrasound literature has shown that measurements can be substantially influenced by operator experience and technique and that inter-rater agreement is typically lower than intra-rater agreement [
3,
4]. This raises concern regarding whether the high diagnostic accuracy observed here would persist if the SHD was measured by less experienced clinicians, and we believe that this also deserves consideration as a limitation.
SHD was measured with the patient supine and the head in a neutral position, whereas mask ventilation was assessed after induction with the head in a “ramped” position. Head and neck positioning have been shown to significantly alter hyomental and skin-to-hyoid distances [
4]. This discrepancy between the position used for measurement and that used for clinical assessment raises the question of whether the derived 1.9-cm cutoff fully reflects the anatomical relationship present under actual clinical conditions. We believe that a discussion of this potential confounder would further strengthen the interpretation of the findings.
We consider that these three points do not diminish the value of the study, but rather highlight directions for future multicenter studies involving multiple operators, which would help confirm the role of SHD in predicting DMV in patients with obesity.
Sincerely,