Response to "Comment on Comparison between the automated TEG®6s and viscoelastic analyzer Haema T4"
Article information
Dear Editor,
We appreciate the authors’ interest in our study [1]. However, several of their criticisms rest on the conflation of distinct study objectives, namely, comparative device analysis, physiological modeling, and clinical validation. As a result, this Letter applies methodological expectations that exceed both the stated aims of our study and common practices in the viscoelastic literature [2].
The principal criticism received concerns the altered citrate-to-sample ratio following dilution and coagulation factor supplementation. This criticism oversimplifies the methodological realities of in vitro coagulation models and viscoelastic testing.
First, all in vitro dilution models inherently deviate from the physiological conditions and cannot fully reproduce the in vivo calcium milieu. This limitation was explicitly acknowledged in our manuscript. The objective of our study was not to recreate a physiological coagulation environment, but rather to compare the analytical behaviors of two viscoelastic platforms under identical experimental conditions.
Second, the criticism applies methodological standards that are neither universally established nor consistently implemented in the viscoelastic testing literature. Numerous published in vitro studies using TEG® and ROTEM® systems have diluted citrated whole blood with saline or colloid solutions without additional citrate supplementation [3,4]. Therefore, our experimental approach is consistent with the methodology established in this field. Furthermore, viscoelastic assays are designed to contain excess calcium to overcome the remaining citrate anticoagulation. Therefore, altered initiation parameters observed during dilution experiments are unlikely to be explained solely by the concentration of citrate vs. calcium and may instead reflect changes in the balance between procoagulant and anticoagulant factors [4,5].
The authors also proposed the use of additional citrate supplementation as a methodological solution. However, this approach introduces non-physiological experimental conditions that may alter the assay behavior. Moreover, supplementation with small volumes of citrate introduces additional technical variability. Therefore, both approaches represent different artificial in vitro conditions and neither can definitively restore physiological coagulation conditions.
Importantly, both viscoelastic systems in our study were exposed to identical sample conditions. Although altered citrate conditions may influence absolute measurements, such effects alone cannot sufficiently explain the systematic and assay-specific differences observed between the two platforms. The assertion that the reported findings merely reflect ‘pre-analytical artifacts’ is speculative and is not supported by the data presented in the Letter. In addition, pre-analytical variability is an unavoidable aspect of routine coagulation testing. Evaluating the robustness and analytical behavior of viscoelastic devices under altered sample conditions may provide clinically relevant information rather than representing a methodological weakness. We agree that the altered citrate-to-blood ratio represents a pre-analytical limitation that may have affected the absolute values. However, because both systems were tested under the same conditions, this limitation alone is unlikely to account for the consistent device-specific differences observed between the TEG®6s and Haema T4, which was the basis of our comparative analysis.
The Letter further argued that a direct comparison between two viscoelastic devices requires external benchmarking against conventional coagulation assays. This criticism rests on a mistaken premise regarding the purpose of comparative device studies. The objective of our investigation was to assess the agreement, interchangeability, and systematic bias between the two viscoelastic technologies and not to validate either platform against plasma-based laboratory assays. These represent fundamentally different scientific questions.
Importantly, viscoelastic testing is not simply a surrogate for prothrombin time, Clauss fibrinogen, or conventional activated clotting time measurements [4]. Conventional coagulation assays and viscoelastic testing assess different aspects of coagulation biology and are known to be incompletely correlated. Therefore, the absence of prothrombin time or Clauss fibrinogen measurements does not invalidate our direct device comparison study because this was not the scientific question under investigation. Viscoelastic testing is a functional whole-blood assessment supported by extensive literature, and has been incorporated into international guidelines. Direct comparisons between viscoelastic platforms are well-established in the literature and are routinely performed without mandatory correlation with conventional coagulation assays or clinical outcomes. The reasoning advanced in the Letter would render a substantial part of the published viscoelastic comparison literature methodologically unacceptable, a conclusion that is not supported by current scientific and clinical practice.
Finally, the criticism that our conclusions require clinical outcome data or transfusion analyses extends beyond the scope of our study. Our manuscript did not claim the superiority of one device over another, nor did it propose novel transfusion algorithms or clinical decision thresholds. Rather, we demonstrated that two commercially available viscoelastic systems produced substantially different results despite superficially similar assay nomenclature and intended clinical applications. This observation is important in itself, as it challenges the assumption that numerical values and treatment thresholds can be transferred interchangeably between platforms.
In summary, although we acknowledge the inherent limitations of all in vitro coagulation models, the methodological considerations raised in the Letter do not invalidate the principal findings of our study. Under identical experimental conditions, two commercially available viscoelastic systems exhibited substantial systematic differences. These observations remain methodologically sound and clinically relevant.