Dear Editor,
We read with great interest the recent article by Nikolaus et al. [
1] comparing the automated TEG
®6s (Haemonetics
®) with the Haema T4 (HT4, Medcaptain
®) viscoelastic analyzer. While the authors aimed to evaluate the agreement between the two devices under various
in vitro conditions, we would like to raise several methodological concerns that substantially limit the interpretability and clinical applicability of the findings.
First, the original investigation is best categorized as an analytical device-comparison study rather than a clinical validation study. Such studies aim to assess agreement between measurement systems under standardized conditions, rather than to establish clinical utility. However, the experimental design of this study fails to meet the fundamental pre-analytical requirements specified for the TEG
®6s. According to the manufacturer’s instructions for use, the TEG
®6s system is strictly validated for citrated whole blood collected at a 1:9 ratio (3.2% sodium citrate to blood) [
2]. In this study, however, the authors added isotonic saline, fibrinogen concentrate, and prothrombin complex concentrate to the initial citrated blood, effectively diluting the citrate-to-blood ratio to approximately 1:19 [
1]. Since the TEG
®6s cartridge dispenses a fixed amount of calcium optimized for a 1:9 ratio, this deviation results in an excess of ionized calcium. Such a calcium imbalance artificially accelerates the coagulation cascade, leading to a non-physiological shortening of the activated clotting time (ACT) and r-time [
3]. Consequently, the reported discrepancies between the two devices and the "hypercoagulable profile" attributed to the HT4 likely reflect pre-analytical artifacts rather than true hemostatic characteristics. Viscoelastic measurements obtained without maintaining the appropriate citrate ratio cannot be interpreted within an established clinical or analytical framework.
Second, this study lacks clinical relevance as it fails to measure standard laboratory parameters—such as Clauss fibrinogen, prothrombin time, or conventional celite-ACT—across the various sample conditions obtained from volunteers. Since these gold-standard values were not correlated with the measurements from the viscoelastic devices, the clinical significance of the findings remains unverified. Existing literature has extensively demonstrated that TEG-derived ACT often correlates poorly with gold-standard ACT devices and traditional coagulation tests due to fundamental differences in their measurement principles [
4,
5]. Consequently, the weak correlation observed between the HT4 and TEG
®6s is an expected outcome when comparing two distinct mechanical methodologies without a stable, standardized reference point [
1]. These discrepancies should not be interpreted as a lack of device performance or clinical utility. Without benchmarking against established clinical metrics, it remains impossible to determine the clinical significance of the numerical differences observed between the two devices.
Third, established research on viscoelastic testing has consistently focused on either (1) correlation with standard laboratory assays or (2) association with clinical outcomes, such as blood loss, transfusion requirements, and thrombotic events [
3,
5]. These evaluations are indispensable for determining whether a device provides clinically actionable information. In contrast, the present study assesses only the numerical discrepancies between two devices [
1], making it impossible to extrapolate these findings to clinical decision-making. Moreover, terms such as “hypercoagulable profile,” “device-specific reference ranges,” and clinical applicability inherently imply a relationship to physiological or clinical states. However, the present study does not establish whether the observed inter-device differences correspond to true alterations in coagulation status or instead reflect methodological or calibration-related biases. For example, higher MA values in one device cannot be directly interpreted as indicating a hypercoagulable state, particularly in the presence of pre-analytical calcium imbalance. To support such interpretations, validation against independent reference standards is required, including correlation with established laboratory assays (e.g., Clauss fibrinogen, platelet function tests) and, ideally, association with clinically relevant outcomes such as bleeding or thrombotic events.
To provide a constructive perspective for evaluating novel viscoelastic devices like the HT4, a structured, multi-phase validation strategy should be adopted in future research. In the initial phase, rigorous analytical validation must be performed in a healthy cohort by strictly adhering to specified pre-analytical conditions to establish dependable baseline reference intervals. Subsequently, ex vivo or clinical samples should be simultaneously benchmarked against gold-standard conventional coagulation assays to identify potential device-specific interferences or artifacts. Ultimately, prospective clinical outcome studies in bleeding patients are required to link novel device parameters with actual clinical endpoints, such as transfusion requirements or surgical bleeding. Such a stepwise approach is essential to ensure both analytical reliability and clinical applicability.
In light of these concerns, the breach of the standard citrate ratio, the lack of comparison with gold-standard laboratory assays, and the absence of correlation with clinical outcomes collectively undermine the validity of the study’s conclusions. To properly evaluate a novel device such as the HT4, it is imperative to adhere to pre-analytical conditions specified by the manufacturer and to include rigorous comparisons with both standard coagulation tests and clinically relevant outcomes. Only by meeting these fundamental scientific criteria can the true analytical performance and clinical utility of the device be appropriately assessed.