Choi: Clinical value of ultrasound-based skin-to-hyoid measurement in obese patients
Traditional indicators such as the Mallampati classification, thyromental distance, and the STOP-Bang score have long been used as standard tools for airway assessment [1]. However, these methods have inherent limitations, including a reliance on subjective judgment and the need for patient cooperation [2]. In obese patients, excessive soft tissue deposition further compromises the predictive performance of these tools [3], and unexpected difficult mask ventilation (DMV) remains a significant clinical concern for anesthesiologists [4].
To address these clinical limitations, ultrasound-based airway assessments have emerged as a promising objective alternative [5]. However, most existing studies have focused primarily on predicting the difficulty of tracheal intubation, and research specifically addressing DMV in obese populations remains limited [6].
The study by Mostafa et al. [7] published in this issue of the Korean Journal of Anesthesiology offers pivotal insights into addressing this clinical gap. In this prospective observational study of 326 patients with a body mass index (BMI) ≥ 35 kg/m2, the authors evaluated the predictive performance of the skin-to-hyoid distance (SHD) for DMV. The SHD demonstrated excellent discriminative ability, with an area under the curve (AUC) of 0.88. Notably, using a cutoff value of 1.9 cm, the SHD showed a high negative predictive value of 99%, suggesting its potential utility as a rule-out tool to identify patients at low risk for DMV. At this threshold, the SHD achieved a sensitivity of 86% and specificity of 83%, substantially outperforming the results reported by Tasdemir et al. [8], who found an AUC of only 0.70 with a lower cutoff of 1.7 cm in a less severely obese cohort.
A key strength of this study is its clinical applicability. The use of handheld ultrasound devices enhances feasibility [7], and the SHD measurement is time-efficient, typically requiring less than a minute and can be performed regardless of the patient’s level of consciousness or cooperation [9]. The authors also suggested that the technique may be easily learned even by relatively inexperienced operators, consistent with the evidence that point-of-care ultrasound psychomotor skills can be acquired with modest training [10]. In addition, compared with previous studies (e.g., Tasdemir et al. [8]), this study included a higher proportion of high-risk patients, with morbidly obese and super-obese individuals accounting for more than 60% of the cohort. The finding that the SHD independently predicted DMV beyond conventional indices such as the STOP-Bang score and Mallampati classification [7] further supports its potential role as a complementary assessment tool, aligning with recent evidence that ultrasound parameters often outperform traditional clinical predictors [11,12].
The DMV incidence of 6.7% observed in this study is consistent with the incidence rates reported in other obese surgical patients, which range from 6.7% to 11.7% [13,14]. This consistency strengthens the external validity of the findings. Furthermore, this study strengthened the robustness of the SHD cutoff value using rigorous statistical methods, including 10 repetitions of stratified 5-fold cross-validation and Firth’s penalized likelihood approach for complete separation cases [7].
Despite these strengths, certain limitations merit consideration in future research. First, although the observed DMV incidence (6.7%) appears adequate for analysis, the single-center design limits generalizability. Second, while the SHD showed promising performance as a stand-alone predictor, it remains unclear whether combining the SHD with established clinical predictors could further improve predictive accuracy. A recent study by De Luis-Cabezón et al. [12] demonstrated that integrating ultrasound parameters with clinical variables into a composite score improved the sensitivity and negative predictive value compared with either approach alone. Future studies evaluating integrated models that incorporate the SHD would be valuable. Third, the optimal cutoff value (1.9 cm) identified in this cohort may not be directly applicable to populations with a lower BMI, and external validation across diverse patient groups is necessary. The discrepancy between the current study’s cutoff value and that of Tasdemir et al. [8] underscores the importance of population-specific validation. Finally, although this study excluded patients with apparent airway abnormalities or suspected difficult laryngoscopy, the SHD’s performance in emergency or unscreened populations remains unclear.
The results from recent ultrasound-based DMV prediction studies provide additional context. Anchalee et al. [15] reported that, while the SHD contributed to a multivariate model predicting severe DMV (AUC 0.89), it was insufficient as a sole predictor in Thai adults. Lin et al. [16] demonstrated that submandibular ultrasound measurements, specifically tongue base thickness, independently predict DMV. These results suggest that different ultrasound approaches may be complementary, and future research should explore whether combining multiple ultrasound parameters (e.g., SHD and tongue thickness) can further improve prediction accuracy.
In conclusion, the study by Mostafa et al. [7] suggests that ultrasound-based SHD measurements may serve as a useful adjunct for airway assessment in obese patients. Its high negative predictive value, ease of acquisition with handheld devices, and independence from patient cooperation make the SHD particularly attractive for preoperative screening. While the findings are encouraging, further validation in multicenter cohorts, integration with existing clinical frameworks, and evaluation of combined ultrasound-clinical prediction models will be essential before widespread adoption in routine practice. As point-of-care ultrasound becomes increasingly accessible [6], SHD measurements represent a promising step toward more objective and reproducible airway risk stratification in this challenging patient population.

Funding

None.

Conflicts of Interest

Yoon Ji Choi has been a member of the editorial board of the Korean Journal of Anesthesiology. There were no other potential conflicts of interest relevant to this article.

References

1. Apfelbaum JL, Hagberg CA, Connis RT, Abdelmalak BB, Agarkar M, Dutton RP, et al. 2022 American Society of Anesthesiologists practice guidelines for management of the difficult airway. Anesthesiology 2022; 136: 31-81.
crossref pmid pdf
2. Roberts S, Cyna AM, Walsh JP, Davis JS. Assessment of anaesthetists’ ability to predict difficulty of bag-mask ventilation. Br J Anaesth 2013; 111: 676-7.
crossref pmid
3. Moon TS, Fox PE, Somasundaram A, Minhajuddin A, Gonzales MX, Pak TJ, et al. The influence of morbid obesity on difficult intubation and difficult mask ventilation. J Anesth 2019; 33: 96-102.
crossref pmid pdf
4. Kheterpal S, Healy D, Aziz MF, Shanks AM, Freundlich RE, Linton F, et al. Incidence, predictors, and outcome of difficult mask ventilation combined with difficult laryngoscopy: a report from the multicenter perioperative outcomes group. Anesthesiology 2013; 119: 1360-9.
crossref pmid
5. Zetlaoui PJ. Ultrasonography for airway management. Anaesth Crit Care Pain Med 2021; 40: 100821.
crossref pmid
6. Petrisor C, Dîrzu D, Trancă S, Hagău N, Bodolea C. Preoperative difficult airway prediction using suprahyoid and infrahyoid ultrasonography derived measurements in anesthesiology. Med Ultrason 2019; 21: 83-8.
crossref pmid pdf
7. Mostafa M, ElGeneidi I, Hasanin A, Ali M, Mostafa H, Naguib NN. Accuracy of ultrasound-measured skin-to-hyoid bone distance for predicting difficult mask ventilation in patients with obesity: a prospective observational study. Korean J Anesthesiol 2026; 79: 352-8.
crossref pmid
8. Tasdemir O, Kocaoglu N, Demir HF, Ugun F, Sagir O. Ultrasound-based airway assessment in obese patients as a valuable tool for predicting difficult airway: an observational study. Braz J Anesthesiol 2024; 74: 844539.
crossref pmid pmc
9. Karakosta A, Pantazi D, Margariti P, Micha G, Samara E, Tzimas P. Training requirements in point-of-care ultrasonography of the upper airway: a feasibility study. J Ultrasound Med 2023; 42: 2513-22.
crossref pmid
10. Breunig M, Chelf C, Kashiwagi D. Point-of-care ultrasound psychomotor learning curves: a systematic review of the literature. J Ultrasound Med 2024; 43: 1363-73.
crossref pmid
11. Wu H, Wang H. Diagnostic efficacy and clinical value of ultrasonography in difficult airway assessment: based on a prospective cohort study. Contrast Media Mol Imaging 2022; 2022: 4706438.
crossref pmid pmc pdf
12. De Luis-Cabezón N, Ly-Liu D, Renedo-Corcostegui P, Santaolalla-Montoya F, Zabala-Lopez de Maturana A, Herrero-Herrero JC, et al. A new score for airway assessment using clinical and ultrasound parameters. Front Med (Lausanne) 2024; 11: 1334595.
crossref pmid pmc
13. Onay M, Kayhan GE, Şanal Baş S, Bilgin M, Kiliç Y, Yelken B, et al. Ultrasonographic evaluation of difficult airway in obese patients: a prospective study. Bariatr Surg Pract Patient Care 2024; 19: 129-34.
crossref pmid pmc pdf
14. Leoni A, Arlati S, Ghisi D, Verwej M, Lugani D, Ghisi P, et al. Difficult mask ventilation in obese patients: analysis of predictive factors. Minerva Anestesiol 2014; 80: 149-57.
pmid
15. Anchalee S, Wasoontrarak K, Benjhawaleemas P, Chatmongkolchart S, Prathep S. Prediction of difficult mask ventilation in Thai adult patients undergoing elective surgery using ultrasound of distance from skin to hyoid bone, and from skin to thyroid isthmus: a prospective cohort observational study. BMC Anesthesiol 2025; 25: 43.
crossref pmid pmc pdf
16. Lin HY, Tzeng IS, Hsieh YL, Kao MC, Huang YC. Submental ultrasound is effective in predicting difficult mask ventilation but not in difficult laryngoscopy. Ultrasound Med Biol 2021; 47: 2243-49.
crossref pmid


ABOUT
ARTICLE CATEGORY

Browse all articles >

BROWSE ARTICLES
AUTHOR INFORMATION
Editorial Office
101-3503, Lotte Castle President, 109 Mapo-daero, Mapo-gu, Seoul 04146, Korea
Tel: +82-2-792-5128    Fax: +82-2-792-4089    E-mail: journal@anesthesia.or.kr                
Business Name: Korean Society of Anesthesiologists
Business Registration: 106-82-07194
Representative: Young-Tae Jeon

Copyright © 2026 by Korean Society of Anesthesiologists.

Developed in M2PI

Close layer
prev next