Korean J Anesthesiol Search

CLOSE


Korean J Anesthesiol > Volume 72(6); 2019 > Article
Tham, Beh, Shariffuddin, and Wang: Unilateral hypoglossal nerve palsy after the use of laryngeal mask airway (LMA) Protector

Abstract

Background

The laryngeal mask airway (LMA) ProtectorTM (Teleflex Medical Co., Ireland) is the latest innovation in the second generation of LMA devices. One distinguishing feature of this device is its integrated, color-coded cuff pressure indicator (Cuff PilotTM technology) which enables continuous cuff pressure monitoring and allows adjustments when necessary; this ensures patient safety due to better monitoring.

Case

We report a case of postoperative unilateral hypoglossal nerve palsy after uncomplicated use of the LMA Protector. To the best of our knowledge, this could be the second reported case.

Conclusions

This case demonstrates that anesthetists need to routinely measure cuff pressure and that the Cuff PilotTM technology is not a panacea for potential cranial nerve injury after airway manipulation.

The laryngeal mask airway (LMA) ProtectorTM (Teleflex Medical Co, Ireland) is the latest innovation in single-use, second generation LMA devices. Similar to previous LMA devices, the LMA ProtectorTM has common features such as a preformed curvature to allow easy insertion [1,2] with bite block, gastric access and high oropharyngeal leak pressure. The airway and cuff are entirely made of silicone, which makes it more flexible and potentially a less traumatic device. It is the only laryngeal mask which combines a pharyngeal chamber and dual gastric drainage channels, designed specifically to channel gastric content away from the airway [2,3]. An important distinguishing feature of the LMA ProtectorTM Airway is its integrated, color-coded (green, yellow, and red zones corresponding to increasing intracuff pressures) cuff pressure indicator (Cuff PilotTM technology) which enables continuous cuff pressure monitoring and allows adjustments when necessary [2,3], thus ensuring patient safety due to better monitoring [4]. We report a case of postoperative unilateral hypoglossal nerve palsy after the uncomplicated use of the LMA ProtectorTM. Written informed consent for the use of images, video and details of the case for publication of this report has been obtained.

Case Report

A 70-year-old man (weight 70 kg, body mass index 25.7 kg/m2) with underlying hypertension and dyslipidemia underwent an elective open hernioplasty for right inguinal hernia under ambulatory surgery. He had normal airway features: good mouth opening, Mallampati score of 1, thyromental distance > 6 cm and normal tongue protrusion. General anesthesia was induced with intravenous propofol 200 mg plus fentanyl 100 μg and a size 4 LMA ProtectorTM Airway was placed smoothly in a single attempt by a senior resident. The cuff was inflated with 25 ml of air and the black line indicator on the cuff pilot valve remained within the green zone throughout the surgery. However, we did not check the intra-cuff pressure using manometry. The oropharyngeal leak pressure was 25 cmH2O. The sternal notch test and bubble test were performed after insertion to confirm the placement of the LMA protector [5]. Anesthesia was maintained with a mixture of sevoflurane and oxygen/air. The patient’s breathing was supported with a pressure support of 8 cmH2O, which generated a tidal volume of 400–450 ml and the maximum minute ventilation attained was 12 L/min with peak pressures of 8–10 cmH2O. He was placed in a supine position with standard American Society of Anesthesiologists monitoring for the surgery which lasted for 180 minutes. The surgery was uneventful, and the patient’s vital signs were stable throughout. Postoperatively, the LMA was removed smoothly when he was awake. Moreover, blood stains or minimal secretions were not observed on the device.
At the post anesthesia care unit, the patient complained of difficulty in chewing food and a weird tongue movement. He had no voice changes or altered taste sensation. On examination, the patient’s tongue was seen to be deviated to the left during active protrusion (Fig. 1). All sensations of the tongue were intact and there were no tongue fasciculations or wasting. The neurological examination revealed no lateralizing signs or limb weakness. The gag and cough reflexes as well as other cranial nerves were normal. The patient was referred to the ENT surgeon the same day. The nasoendoscopy examination was unremarkable. The working diagnosis was that of an isolated left hypoglossal nerve palsy or neuropraxia. He was allowed to go home the same day with reassurance, oral prednisolone for one week, and instructed for follow up at the ENT outpatient clinic. Neuroimaging was not required. He achieved complete recovery 3 months after the injury (Fig. 1, Supplementary Video 1) and was subsequently discharged from the follow-up clinic.

Discussion

Hypoglossal nerve injury is a rare but distressing complication of airway management that arises during general anesthesia [6]. Although there have been reports of its occurrence after the use of other types of LMA [7-11], our case involves the use of the novel LMA ProtectorTM Airway. To the best of our knowledge, this could be the second reported case. The first was reported by Leong et al. in a poster presented at the European Anesthesiology Conference 2018 in Copenhagen (poster title: unilateral hypoglossal nerve palsy after the use of a novel supraglottic airway device; 01AP03-8).
The hypoglossal nerve innervates all the extrinsic and intrinsic muscles of the tongue, except the palatoglossus which is innervated by the vagus nerve. An injury to the hypoglossal nerve causes ipsilateral tongue deviation (pathognomonic), with dysarthria and dysphagia in severe cases. The tongue deviates towards the side that is affected due to the unopposed action of the contralateral genioglossus [12]. We postulate that our patient’s left hypoglossal nerve was compressed by the distended LMA cuff against surrounding structures such as the hyoid bone. The hypoglossal nerve exits the cranium via the hypoglossal canal, and travels alongside the internal and external carotid arteries, before passing above the hyoid bone to innervate the tongue muscles. The nerve becomes superficial at the level of the angle of the mandible, passing just above the greater horn of the hyoid to enter the mouth. This is a potential site for compression injury [6,12] and male patients may be more susceptible due to the presence of larger hyoid bones. Some authors have suggested that the LMA cuff insufflation may be associated with hypoglossal nerve injury. In these reported cases, the cuff insufflation volumes were 15 to 40 ml but intraoperative cuff pressure monitoring or titration has not been mentioned in these reports [7,8,10]. Although the recommended cuff insufflation volume varies according to the size and type of LMA used, excessive pressure in the volume cuff, especially during longer cases, exacerbated by the use of nitrous oxide (N2O) may cause injury due to the malposition of the airway devices [7,8,10,11]. Some authors have theorized that intermittent routine cuff pressure monitoring could decrease the incidence of hypoglossal nerve palsy [6,13]. The position of a patient changes after securing the airway while turning from a supine position during surgical preparation and draping, and this can cause cuff malposition and predispose the patient to hypoglossal nerve trauma. Therefore, it is recommended to routinely perform intermittent cuff pressure monitoring especially during long surgeries and when N2O is administered. We should also check the patient’s position intermittently with special focus on the head and airway securement [6].
Hypoglossal nerve injury is typically diagnosed postoperatively after a thorough workup to exclude stroke, endotracheal trauma, airway hematoma, or impending airway obstruction [6]. The symptoms and signs of hypoglossal neurapraxia are often self-limiting and 43% of diagnosed patients achieve resolution within 6 weeks of surgery and an additional 40% are symptom free within 6 months after surgery [6]. Therefore, treatment is usually supportive and follow up is done until resolution of symptoms and signs. A short course of steroid therapy is used to decrease the swelling, especially in cases with suspected airway edema after manipulation. However, there are no controlled studies on the benefits of this treatment for hypoglossal nerve neuropraxia.
Our patient had hypoglossal nerve injury in spite of using the LMA ProtectorTM Airway with Cuff PilotTM technology. The possible contributing factors include cuff over-inflation with failure of the cuff pressure indicator and inappropriate placement of the device. In this case, the LMA protector was inserted smoothly in the first attempt and correct placement was confirmed by with an oropharyngeal leak pressure at 25 cmH2O; therefore, inappropriate placement of the device was unlikely to be the cause. We have some experience with the use of this device for airway management in obese patients and recently presented a poster on this topic at the KoreAnesthesia, 2018 in Seoul (poster title: evaluation of the clinical performance of LMA Protector in the moderately obese patients; ABST-000411). None of the obese patients in the study suffered from hypoglossal nerve palsy. Primary research assessing the effects of LMA-ProtectorTM use is sparse [2]. There have been 2 studies which evaluated the use of the LMA Protector in normal population which did not report any serious complication [1,14].
Although the Cuff PilotTM technology is purported to allow continuous cuff pressure monitoring at a glance, anesthetists should remain vigilant with intermittent monitoring of cuff pressures using manometry. This rare isolated case of hypoglossal nerve injury associated with the use of the novel LMA ProtectorTM Airway suggests that its unique in-built Cuff PilotTM technology is not a panacea for potential cranial nerve injury after airway manipulation. Anesthetists should practice routine cuff pressure monitoring with manometry and careful selection of supraglottic airway devices for short procedures. The size of the device may be decreased in spontaneously breathing patients because this may limit pharyngolaryngeal morbidities such as sore throat and hoarseness of voice [15]. The manufacturer of the LMA-ProtectorTM recommended sizes 3, 4, and 5 for patients weighting 30–50 kg, 50–70 kg, and 70–100 kg respectively. All 3 airway sizes have a similar maximum intra-cuff pressure of 60 cmH2O [3].

Conflicts of Interest

No potential conflict of interest relevant to this article was reported.

Authors’ contribution

Li Yeen Tham (Data curation; Investigation; Writing–original draft; Writing–review & editing)

Zhi Yuen Beh (Conceptualization; Resources; Validation; Writing–original draft; Writing–review & editing)

Ina Ismiarti Shariffuddin (Conceptualization; Resources; Supervision; Writing–original draft; Writing–review & editing)

Chew Yin Wang (Supervision; Validation; Writing–original draft; Writing–review & editing)

Supplementary Materials

Futher detailes are presented in the online version of this article (Available from https://doi.org/10.4097/kja.d.18.00354).

Fig. 1.
Patient with left hypoglossal nerve palsy after the use of laryngeal mask airway protector. (A) Left tongue deviation after the surgery. (B) Complete recovery 3 months after the injury.
kja-d-18-00354f1.jpg

References

1. Sng BL, Ithnin FB, Mathur D, Lew E, Han NR, Sia AT. A preliminary assessment of the LMA protectorTM in non-paralysed patients. BMC Anesthesiol 2017; 17: 26.
crossref pmid pmc
2. Van Zundert AA, Skinner MW, Van Zundert TC, Luney SR, Pandit JJ. Value of knowing physical characteristics of the airway device before using it. Br J Anaesth 2016; 117: 12-6.
crossref pmid pmc pdf
3. LMA Protector Airway [Internet]. Westmeath, Ireland: Teleflex; 2017 [cited 2018 Dec 14]. Available from https://www.teleflex.com/usa/en/product-areas/anesthesia/airway-management/lma-airways/product-literature/AM_SUP_LMA-Protector-Brochure_BR_MC-001649.pdf

4. Wong DT, Tam AD, Mehta V, Raveendran R, Riad W, Chung FF. New supraglottic airway with built-in pressure indicator decreases postoperative pharyngolaryngeal symptoms: a randomized controlled trial. Can J Anaesth 2013; 60: 1197-203.
crossref pmid pdf
5. Timmermann A, Bergner UA, Russo SG. Laryngeal mask airway indications: new frontiers for second-generation supraglottic airways. Curr Opin Anaesthesiol 2015; 28: 717-26.
crossref pmid
6. Shah AC, Barnes C, Spiekerman CF, Bollag LA. Hypoglossal nerve palsy after airway management for general anesthesia: an analysis of 69 patients. Anesth Analg 2015; 120: 105-20.
crossref pmid pmc
7. Nagai K, Sakuramoto C, Goto F. Unilateral hypoglossal nerve paralysis following the use of the laryngeal mask airway. Anaesthesia 1994; 49: 603-4.
crossref pmid
8. King C, Street MK. Twelfth cranial nerve paralysis following use of a laryngeal mask airway. Anaesthesia 1994; 49: 786-7.
crossref pmid
9. Umapathy N, Eliathamby TG, Timms MS. Paralysis of the hypoglossal and pharyngeal branches of the vagus nerve after use of a LMA and ETT. Br J Anaesth 2001; 87: 322.
pmid
10. Stewart A, Lindsay WA. Bilateral hypoglossal nerve injury following the use of the laryngeal mask airway. Anaesthesia 2002; 57: 264-5.
crossref pmid
11. Trümpelmann P, Cook T. Unilateral hypoglossal nerve injury following the use of a ProSeal laryngeal mask. Anaesthesia 2005; 60: 101-2.
crossref pmid
12. Walker HK. Cranial Nerve XII: The Hypoglossal Nerve. In: Clinical Methods: The History, Physical, and Laboratory Examinations. 3rd ed. Edited by Walker HK Hall WD Hurst JW: Boston, Butterworths. 1990.

13. Seet E, Yousaf F, Gupta S, Subramanyam R, Wong DT, Chung F. Use of manometry for laryngeal mask airway reduces postoperative pharyngolaryngeal adverse events: a prospective, randomized trial. Anesthesiology 2010; 112: 652-7.
crossref pmid
14. Tan LZ, Tan DJ, Seet E. Laryngeal mask airway protectorTM: Advanced uses for laparoscopic cholecystectomies. Indian J Anaesth 2017; 61: 673-5.
crossref pmid pmc
15. Grady DM, McHardy F, Wong J, Jin F, Tong D, Chung F. Pharyngolaryngeal morbidity with the laryngeal mask airway in spontaneously breathing patients: does size matter? Anesthesiology 2001; 94: 760-6.
crossref pmid


ABOUT
ARTICLE CATEGORY

Browse all articles >

BROWSE ARTICLES
AUTHOR INFORMATION
Editorial Office
101-3503, Lotte Castle President, 109 Mapo-daero, Mapo-gu, Seoul 04146, Korea
(서울특별시 마포구 마포대로 109 롯데캐슬 프레지던트 101동 3503호)
Tel: +82-2-792-5128    Fax: +82-2-792-4089    E-mail: journal@anesthesia.or.kr                
Business Name: Korean Society of Anesthesiologists (대한마취통증의학회)
Business Registration: 106-82-07194
Representative: Jun Heum Yon (연준흠)

Copyright © 2024 by Korean Society of Anesthesiologists.

Developed in M2PI

Close layer
prev next