Korean J Anesthesiol Search

CLOSE


Korean Journal of Anesthesiology 2008;54(5):493-500.
DOI: https://doi.org/10.4097/kjae.2008.54.5.493   
Agreement Evaluation between Noninvasive Oscillometric Blood Pressure Measured by the Tram 800SL Module of Solar 8000M(R) with Invasive Arterial Blood Pressure Measured by the AMK 250 AutoTransducer(R) during General Anesthesia.
Kwan Sik Park, Sang Kee Min, Sook Young Lee, Jin Soo Kim, Jong Yeop Kim, Sung Yong Park, Young Pyo Son
Department of Anesthesiology and Pain Medicine, Ajou University School of Medicine, Suwon, Korea. anesmin@nate.com
Abstract
BACKGROUND
Blood pressure (BP) varies considerably during general anesthesia. Accurate BP measurement is critical for appropriate treatment, especially during hypotension and hypertension. Here we evaluated whether the noninvasive oscillometric BP measurement technique accurately reflects BP measured by the direct intraarterial technique.
METHODS
A total of 256 samples were extracted from 10 patients operated on under general anesthesia. Systolic, diastolic and mean BP were analyzed according to the level of BP; hypotension, normotension, and hypertension. Repeatability of the noninvasive BP measurement were analyzed with repeatability coefficients and percent errors. Agreement between the two BP measurements were analyzed with a Bland-Altman and Modified Bland-Altman analysis.
RESULTS
The repeatability coefficient for mean BP of the noninvasive oscillometric BP measurement was 6.34. Percent errors of mean BP were smaller than those of systolic and diastolic BP. All biases were less than 5 mmHg at all BP levels. Most BP agreements were larger than 8 mmHg except all-range mean BP (7.7), hypotensive diastolic BP (6.7), and mean BP (6.2). During hypotension, only mean BP was within the 95% of confidence interval (CI) of bias and limit of agreement. During hypertension, only systolic BP could meet the criteria.
CONCLUSIONS
Only mean BP during hypotension and systolic BP during hypertension accurately reflect the direct intraarterial BP. But, diastolic BP does not reflect the direct intraarterial BP over all BPs.
Key Words: accuracy; agreement; bias; measurement; precision; repeatability
TOOLS
Share :
Facebook Twitter Linked In Line it
METRICS Graph View
  • 0 Crossref
  •    
  • 1,456 View
  • 8 Download


ABOUT
ARTICLE CATEGORY

Browse all articles >

BROWSE ARTICLES
AUTHOR INFORMATION
Editorial Office
101-3503, Lotte Castle President, 109 Mapo-daero, Mapo-gu, Seoul 04146, Korea
Tel: +82-2-792-5128    Fax: +82-2-792-4089    E-mail: journal@anesthesia.or.kr                

Copyright © 2024 by Korean Society of Anesthesiologists.

Developed in M2PI

Close layer
prev next