Ventilation Impairment due to Obstruction of a Bifurcated Connector: A case report.
Deok Hee Lee, Young Jun Chang
Department of Anesthesiology and Pain Medicine, College of Medicine, Yeungnam University, Daegu, Korea. dhlee415@med.yu.ac.kr
Abstract
Ventilation impairment sometimes developes in patients with a double-lumen tube (DLT) during thoracic anesthesia. A 27-year old man who suffered from a left pneumothorax showed ventilation impairment with right lung obstruction after the induction of anesthesia. We checked breathing sounds before and after clamping each lumen. Correct positioning was confirmed by direct visualization using a flexible fiberoptic bronchoscope and by chest X-ray. The DLT was removed and replaced by a single lumen endotracheal tube because of severe ventilation impairment. On careful examination of the DLT, manufacturing defects in a bifurcated connector were noted, and its openings were occluded with a thin plastic membrane. After replacing the DLT, the ventilation impairment was improved. It is important to test DLT prior to use, and to maintain an awareness that airway obstruction can occur due to structural defects.
Key Words: bifurcated connector; double-lumen tube; obstruction; ventilation impairment


ABOUT
ARTICLE CATEGORY

Browse all articles >

BROWSE ARTICLES
AUTHOR INFORMATION
Editorial Office
101-3503, Lotte Castle President, 109 Mapo-daero, Mapo-gu, Seoul 04146, Korea
Tel: +82-2-792-5128    Fax: +82-2-792-4089    E-mail: journal@anesthesia.or.kr                

Copyright © 2024 by Korean Society of Anesthesiologists.

Developed in M2PI

Close layer
prev next