Inaccurate Reading of Pulse Oximeter Due to Methemoglobinemia: A case report.
Kyung Hee Kim, Hae Sung Choi
Department of Anesthesiology, Taejeon St. Mary's Hospital, The Catholic University of Korea, Taejeon, Korea.
Abstract
Pulse oximeters measure the arterial oxygenation by determining the color of the blood between a light source and a photodetector. The light source consists of two light-emitting diodes (LEDs) that emit light at known wave lengths, 660 nm red light and 940 nm infrared light. The ratio of pulse-added red absorbance at 660 nm to pulse-added infrared absorbance at 940 nm is used to generate the oximeter's estimate of arterial saturation (SpO2). Pulse oximeters can determine the concentration of only two hemoglobins, reduced (HHb) and oxyhemoglobin (HbO2), so they can't distinguish the dyshemoglobins (methemoglobin, carboxyhemoglobin) which have light absorbances similar to that of HHb or HbO2. If the concentration of dyshemoglobin is above the normal range, pulse oximeters would give erroneous SpO2 readings. We experienced a case which showed a low SpO2 reading but had normal ABGA findings due to unsuspected methemoglobinemia.
Key Words: hemoglobin, methemoglobin; Monitoring, pulse oximetry; Oxygen, saturation


ABOUT
ARTICLE CATEGORY

Browse all articles >

BROWSE ARTICLES
AUTHOR INFORMATION
Editorial Office
101-3503, Lotte Castle President, 109 Mapo-daero, Mapo-gu, Seoul 04146, Korea
Tel: +82-2-792-5128    Fax: +82-2-792-4089    E-mail: journal@anesthesia.or.kr                

Copyright © 2024 by Korean Society of Anesthesiologists.

Developed in M2PI

Close layer
prev next