Korean J Anesthesiol Search

CLOSE


Korean Journal of Anesthesiology 1998;35(5):952-957.
DOI: https://doi.org/10.4097/kjae.1998.35.5.952   
Incidence of Malposition and it's Affecting Factors of Left-Sided Double-Lumen Endotracheal Tube.
Choon Soo Lee, Chong Kweon Chung, Jeong Uk Han, Hong Sik Lee, Tae Jung Kim, Young Deog Cha, Hong Lee
Abstract
Backgroud: In one-lung ventilation using the left-sided double-lumen tube (LDLT), it is important to place the LDLT in correct position to maintain adequate ventilation. We investigated the frequency of and the factors affecting the LDLT malposition in endotracheal intubation.
Methods
Ninety one (55 male and 36 female) patients were observed. After endotracheal intubation, using 35 and 37 Fr. sized Robertshaw type LDLT, auscultation and fiberoptic bronchoscope were performed to make sure the correct position of LDLT. The heights, weights, age, and sex were noted. The lengths and diameters of trachea, and the angles and diameters of both bronchi on chest x-ray were measured for comparison.
Results
Normal in auscultation and gross malposition were 87.9% and 12.1%. Among those normal in auscultation, normal in bronchoscope, advancing and removing fine malposition were 66.2%, 18.8% and 15.0%, respectively. The angle of left bronchus is 37.71+/-4.60degrees in normal in ausculation and 37.71+/-4.60degrees in gross malposition. The length of trachea is 13.41+/-0.90 cm in normal in bronchoscope, 14.49+/-0.78 cm in advancing fine malposition and 11.86+/-0.35 cm in removing fine malposition. The patient's height is 167.27+/-7.12 cm in normal in brochoscope, 172.45+/-6.67 cm in advancing fine malposition and 163.12+/-6.54 cm in removing fine malposition.
Conclusions
The angle of left bronchus is a factor affecting gross malposition. And the length of trachea and the patient's height are factors affecting fine malposition. Thus it is necessary to obtain in advance information on patient's height, length of trachea and angle of left bronchus on chest x-ray, to reduce the occurrence of the LDLT malposition.
Key Words: Equipment: tubes; left-sided double-lumen endotracheal tube; Ventilation: one-lung
TOOLS
Share :
Facebook Twitter Linked In Line it
METRICS Graph View
  • 0 Crossref
  •    
  • 2,412 View
  • 2 Download


ABOUT
ARTICLE CATEGORY

Browse all articles >

BROWSE ARTICLES
AUTHOR INFORMATION
Editorial Office
101-3503, Lotte Castle President, 109 Mapo-daero, Mapo-gu, Seoul 04146, Korea
Tel: +82-2-792-5128    Fax: +82-2-792-4089    E-mail: journal@anesthesia.or.kr                
Business Name: Korean Society of Anesthesiologists
Business Registration: 106-82-07194
Representative: Young-Tae Jeon

Copyright © 2026 by Korean Society of Anesthesiologists.

Developed in M2PI

Close layer
prev next