Korean J Anesthesiol Search

CLOSE


Korean Journal of Anesthesiology 1992;25(2):433-435.
DOI: https://doi.org/10.4097/kjae.1992.25.2.433   
Occurrence of Hypercarbia due to a Defective Unidirectional Valve.
Eun Ha Kim, Tae Ho Chung
Department of Anesthesiology, National Police Hospital, Seoul, Korea.
Abstract
Hypercarbia can occur in spite of adequate ventilation if removal of carbon dioxide from the inspired gases is inadequate. Neither blood pressure nor heart rate is a reliable indicator of hypercarbia. A 53-year old male, ASA physical status II with mild hypertensive patient underwent primary repair because of rotator cuff tear of the shoulder. Anesthesia was induced with thiopental-succinylcholine and maintained with pancuronium- Halothane-N2O-O2 in a semiclosed circle system. We report occurrence of hypercarbia due to replacement with an incorrect unidirectional valve which is too small to function properly.
Key Words: Hypercarbia; Unidirectional valve; Semiclosed circle system