BACKGROUND
Currently available warming devices are often heavy and cumbersome, requiring development of more portable, user friendly, high efficiency fluid and blood warmer. The intravenous fluid heating capabilities of a new, heat-plate print circuit board (PCB) based warmer at various low flow rates were tested.
METHODS
Model Joyother BM-1(R) was investigated for normal saline (0.9% NaCl) and colloid at various infusion rates (60-6000 ml/hr). Final temperatures were measured by electronic thermometer after passing through the warmer. Effective warming was defined as a fluid temperature > or = 32degrees C. Ambient temperature was maintained at 22-25degrees C. Degree of heating capability and temperature decrease were compared and correlated between different flow rates.
RESULTS
The device warmed the room temperature crystalloid and colloid efficiently. Its warming capability was continuously improved as the flow rates increased in all tested flow rates, reaching maximum 41degrees C. After the warming, the temperature decrease showed high negative correlation with the flow rates. The extent of cooling was less in colloid. No overheating was noted at sudden brake.
CONCLUSIONS
Joyother BM-1(R) heated crystalloid and colloid sufficiently and safely enough for clinical application (exit temperature > or = 35degrees C) at various flow rates. The warming capacity and the length of the intravenous tube infusion system determined the efficiency of the warmer. Further study is warranted for the efficiency of warming for the blood and its element changes.
Key Words: hypothermia; intravenous fluids; warmer
TOOLS
Share :
Facebook Twitter Linked In Line it
METRICS Graph View
  • 2 Crossref
  •    
  • 1,607 View
  • 20 Download


ABOUT
ARTICLE CATEGORY

Browse all articles >

BROWSE ARTICLES
AUTHOR INFORMATION
Editorial Office
101-3503, Lotte Castle President, 109 Mapo-daero, Mapo-gu, Seoul 04146, Korea
Tel: +82-2-792-5128    Fax: +82-2-792-4089    E-mail: journal@anesthesia.or.kr                

Copyright © 2024 by Korean Society of Anesthesiologists.

Developed in M2PI

Close layer
prev next