Wong and Lim: Response to comments on “Awake supraglottic airway guided flexible bronchoscopic intubation in patients with anticipated difficult airways: a case series and narrative review”
We thank the authors for their reply; many of their initial statements were covered in our article. We agree that the cuff inflation (to 60 cmH2O) of a second-generation supraglottic airway device (SAD) allows for better airway seal. However, we keep the cuff deflated for various reasons: It minimizes the oropharyngeal pressure and may help relieve patient discomfort; it avoids the risk of obscuring the glottic view; and the cuff can be inflated later, if required, for instance, for positive pressure ventilation. In one study, with no cuff inflation, the optimal glottic view (only cords seen) occurred in 30% of patients compared to the 0% occurring with an intracuff volume of 40 ml [1].
Oxygen insufflation at 10–15 L/min using our described technique is highly unlikely to cause gastric insufflation. Firstly, oxygen insufflation at the entrance of the proximal end of the SAD ventilation port can be considered to be in an open circuit. Any buildup of pressure would dissipate quickly via the path of least resistance (retrograde to the atmosphere) rather than moving anterograde through the ventilation port occupied by the bronchoscope. Secondly, 50 L/min of high-flow nasal oxygen only achieves a mean airway pressure of up to 7 cmH2O [2], which would not overcome a resting lower esophageal sphincter pressure of 15–35 mmHg (20–47 cmH2O) [3] to cause gastric insufflation. Awake insertion of SADs does not alter the gastroesophageal barrier pressure [4]. The following two studies evaluating gastric insufflation with antral ultrasonography also supported a low risk of gastric insufflation. First, face mask ventilation at 10 cmH2O under anesthesia without neuromuscular blockade showed that gastric insufflation occurred in 0% of cases [5]. Second, all 30 fasted patients who were administered with high-flow nasal oxygen at 70 L/min for a total of 30 min had grade 0 or 1 antrum scores, consistent with fasting. Moreover, in the sitting position, the stomach is lower in relation to the glottis; therefore, the risk of regurgitation and aspiration is potentially reduced.

NOTES

No potential conflict of interest relevant to this article was reported.

Author Contributions

Patrick Wong (Conceptualization; Supervision; Validation; Writing – original draft; Writing – review & editing)

Wan Yen Lim (Conceptualization; Resources; Validation; Writing – original draft; Writing – review & editing)

References

1. Keller C, Puhringer F, Brimacombe JR. Influence of cuff volume on oropharyngeal leak pressure and fibreoptic position with the laryngeal mask airway. Br J Anaesth 1998; 81: 186-7.
crossref pmid pdf
2. Ritchie JE, Williams AB, Gerard C, Hockey H. Evaluation of a humidified nasal high-flow oxygen system, using oxygraphy, capnography and measurement of upper airway pressures. Anaesth Intensive Care 2011; 39: 1103-10.
crossref pmid
3. Keller C, Brimacombe J. Resting esophageal sphincter pressures and deglutition frequency in awake subjects after oropharyngeal topical anesthesia and laryngeal mask device insertion. Anesth Analg 2001; 93: 226-9.
crossref pmid
4. Bouvet L, Albert ML, Augris C, Boselli E, Ecochard R, Rabilloud M, et al. Real-time detection of gastric insufflation related to facemask pressure--controlled ventilation using ultrasonography of the antrum and epigastric auscultation in nonparalyzed patients: a prospective, randomized, double-blind study. Anesthesiology 2014; 120: 326-34.
crossref pmid
5. Lam KK, Giron-Arango L, Behringer EC, Conte AH, Chan VW, Perlas A. No evidence of gastric distension after 30 min of heated and humidified high-flow nasal oxygen in fasting subjects. Br J Anaesth 2019; 123: e451-2.
crossref


ABOUT
ARTICLE CATEGORY

Browse all articles >

BROWSE ARTICLES
AUTHOR INFORMATION
Editorial Office
101-3503, Lotte Castle President, 109 Mapo-daero, Mapo-gu, Seoul 04146, Korea
Tel: +82-2-792-5128    Fax: +82-2-792-4089    E-mail: journal@anesthesia.or.kr                

Copyright © 2024 by Korean Society of Anesthesiologists.

Developed in M2PI

Close layer
prev next