Korean Journal of Anesthesiology 1998;35(6):1119-1123.
DOI: https://doi.org/10.4097/kjae.1998.35.6.1119   
A Comparison of Hemodynamic Indices Derived by Invasive Monitoring and Transesophageal Echocardiography in Patients with Aortic Stenosis.
Eun Sook Yoo, Young Lan Kwak, Sang Beom Nam, Choon Soo Lee, Dong Woo Han, Sang Gun Han, Young Seok Lee, Yong Woo Hong
1Department of Anesthesiology, Ajou University College of Medicine, Suwon, Korea.
2Department of Anesthesiology, Yonsei University College of Medicine, Seoul, Korea.
3Department of Anesthesiology, Inha University College of Medicine, Inchon, Korea.
Abstract
BACKGROUND
Intraoperative fluid management should be aimed at maintaining appropriate left-sided filling pressures. The pulmonary capillary wedge pressure (PCWP) will overestimate the left ventricular end-diastolic volume (LVEDV) when ventricular compliance is markedly reduced in patients with aortic stenosis. Intraoperative transesophageal echocardiography (TEE) is useful for monitoring global left ventricular function and change of preload. This study was undertaken to evaluate preload derived by conventional invasive monitoring technique compare with preload obtained simultaneously from TEE in patients with aortic stenosis.
METHODS
Fifteen patients with aortic stenosis who underwent aortic valve replacement were examined. The preload was examined by the short axis view of left ventricle with TEE at the level of the papillary muscles. For each patient, simultaneous measurements of PCWP, thermodilution cardiac output and left ventricular end-diastolic area (LVEDA) measured by TEE were made after the induction of anesthesia and after surgery.
RESULTS
The correlation between echo-derived LVEDA and thermodilution cardiac index (CI) (r=0.53, p<0.05) or stroke index (SI) (r=0.56, p<0.05) was good after surgery, but the correlation was not found after induction of anesthesia. No correlation was observed between PCWP and CI or SI.
CONCLUSIONS
The PCWP did not provide a reliable estimate of preload and did not allow good prediction of cardiac index. These findings demonstrate that, in patients with aortic stenosis who underwent aortic valve replacement, TEE provides a better index of left ventricular preload than conventional invasive hemodynamic monitoring particularly after surgery.
Key Words: Anesthesia, cardiac; Monitoring, pulmonary capillary wedge pressure, transesophageal echocardiography


ABOUT
ARTICLE CATEGORY

Browse all articles >

BROWSE ARTICLES
AUTHOR INFORMATION
Editorial Office
101-3503, Lotte Castle President, 109 Mapo-daero, Mapo-gu, Seoul 04146, Korea
Tel: +82-2-792-5128    Fax: +82-2-792-4089    E-mail: journal@anesthesia.or.kr                

Copyright © 2024 by Korean Society of Anesthesiologists.

Developed in M2PI

Close layer
prev next